19 Air buoyancy correction ("BEST")
■
63
En
gl
is
h
19.4.1 Weighing weights
Once you have accepted the current air density by pressing «
OK
», the balance can be used as in normal
weighing mode, i.e. taring with «
T
» etc., for example.
For marking purposes, the weight display contains the weight reading, which has been offset against the air
buoyancy correction; preferred readings are also marked with a small circle ("o"). This circle is included in
the printout. The value of the material density, which is used for the correction, is specified in the info line.
19.4.2 Changing the air density
Press «
F
» briefly twice to confirm the air density.
19.4.3 Changing the material density
O
+
1,94683
9
Current weight, "air buoyancy-corrected"
8.000000 9.ccm
Display
Key
Step
1,200000
«
C
»
Switch to enter the new air density
OK
SET
- - - - - - -
«
E
»
Start the air density entry
AIR DENSITY
1.200000
- - - - - - -
«
A
»,«
B
»
«
C
»,«
D
»
Enter the new air density
AIR DENSITY
1.198000
- - - - - - -
«
E
»
Confirm the entry
AIR DENSITY
1.198000
1,198000
«
esc
»
Go back to the air density confirmation
OK
SET
Display
Key
Step
O
+
1,94683
9
«
A
»
Switch to enter the new material density
8.000000 9.cCm
- - - - - - -
«
E
»
Start entering the new material density
MAT. DENSITY 8.000000
- - - - - - -
«
A
»,«
B
»
«
C
»,«
D
»
Enter the new material density
MAT. DENSITY 8.123456
- - - - - - -
«
E
»
Confirm the entry
MAT. DENSITY 8.123456
O
+
1,94683
9
«
esc
»
Go back to weighing with air buoyancy correction
8.123456 9.CCM