Page 43
V 1.3 – February 2016
U s e r I n s t r u c t i o n M a n u a l
Set Remote Method
PC Sends
S!
stage1startpos,stage1endpos,stage1time,stage1temperature,stage1flow,stage2startp
os,stage2endpos,stage2time,stage2temperature,stage2flow,stage3startpos,stage3en
dpos,stage3time,stage3temperature,stage3flow,stage4startpos,stage4endpos,stage4t
ime,stage4temperature,stage4flow,stage5startpos,stage5endpos,stage5time,stage5te
mperature,stage5flow
Where values are same as above.
Query Local Method
PC Sends Mxx?<cr>
Mistral replies
Mxx!
methodname,stage1startpos,stage1endpos,stage1time,stage1temperature,stage1flow
,stage2startpos,stage2endpos,stage2time,stage2temperature,stage2flow,stage3startp
os,stage3endpos,stage3time,stage3temperature,stage3flow,stage4startpos,stage4en
dpos,stage4time,stage4temperature,stage4flow,stage5startpos,stage5endpos,stage5t
ime,stage5temperature,stage5flow
Where values are as above plus:
xx is method number 0-14
methodname is ascii string of no more than 15 characters
Set Local Method
PC Sends
Mxx!
methodname,stage1startpos,stage1endpos,stage1time,stage1temperature,stage1flow
,stage2startpos,stage2endpos,stage2time,stage2temperature,stage2flow,stage3startp
os,stage3endpos,stage3time,stage3temperature,stage3flow,stage4startpos,stage4en
dpos,stage4time,stage4temperature,stage4flow,stage5startpos,stage5endpos,stage5t
ime,stage5temperature,stage5flow<cr>
If a command is successful the Mistral will reply ok, if not the Mistral will reply errr.
Remote Communications Protocol