background image

110-031 PSC/EN

5

LCD Display

Keypad

High Limit Warning 

Indicator

Flashes if you 

attempt to enter a 

set point higher than 

the High Limit value. 

Lit steady when High 

Limit value is being 

set 

Heating Symbol

Lit when heater is 

active

Fault / Warning 

Symbol

Lit when fault or 

warning condition is 

detected

Set Point Indicator

Flashes when 

set point is being 

changed; lit steady 

during normal 

operation

Timer Symbol

Lit when timer 

is being set or 

is in use

Over-Temperature / 

Low Liquid Symbol

Lit when over-

temperature or low 

liquid level fault is 

detected

Calibration Mode 

Indicators

Lit when unit 

calibration offset is 

being set

Version Indicator

Lit when firmware 

version ID is being 

displayed on the 

lower readout

Temperature Unit 

Symbol

Shows 

temperature unit in 

use when lit

Upper Readout

Displays actual 

bath temperature

Lower Readout

Displays set point 

temperature 

during normal 

operation; displays 

operational 

information when 

Menu button is 

pressed

Temperature 

Resets Indicator

Lit when preset 

is being set or is 

in use

Loss of Power 

Indicator

Lit when electrical 

power is restored 

after power disruption

Power Button

Turns power to 

the Controller On 

and Off

Timer Button

Used to start 

and stop the 

timer

Selection Buttons

Used to increase 

or decrease 

temperature 

settings and 

operational 

parameters

Temperature 

Preset Buttons

Used to select 

frequently used 

user-defined 

set point 

temperatures

Menu Button

Used to select and set various 

operational parameters

www.calcert.com

[email protected]

1.888.610.7664

0

5

10

15

20

25

30

Summary of Contents for WBE10A11B

Page 1: ...110 031 PSC EN 1 110 031 PSC EN 25 February 2014 Operator s Manual General Purpose Water Baths www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30...

Page 2: ...h ON 10 Adjusting the Set Point Temperature 10 Setting and Using the Timer 11 Using the Temperature Presets 11 Entering a Calibration Offset 12 Resetting the Factory Default Values 13 Loss of Power Re...

Page 3: ...according to the directions in this manual and common safety procedures your Water Bath should provide safe and reliable temperature control Please ensure that all individuals involved in the install...

Page 4: ...luid in the reservoir Periodically check the reservoir to ensure that the liquid depth is within acceptable levels Always refill the reservoir using the same bath fluid type that is already in the res...

Page 5: ...packing Your Water Bath Your water bath is shipped in a special carton Retain the carton and all packing materials until the unit is completely assembled and working properly Set up and run the unit i...

Page 6: ...ID is being displayed on the lower readout Temperature Unit Symbol Shows temperature unit in use when lit Upper Readout Displays actual bath temperature Lower Readout Displays set point temperature d...

Page 7: ...ss than 2 inches 5 08 cm from the bottom of the unit Operation of the bath without fluid will not damage the heater but will cause permanent discoloration of the tank and will not provide accurate tem...

Page 8: ...ure to follow your organization s procedures and practices regarding the safe lifting and relocation of heavy objects WARNING To avoid the potential for burns allow the Water Bath to cool completely b...

Page 9: ...he unit The extension cord must not cause more than a 10 drop in voltage to the unit Controller Setup Power On 1 Place the Mains Switch on the rear of the unit in the On position The three decimal poi...

Page 10: ...er readout 2 Use the and buttons to adjust the set point displayed on the lower readout to the desired temperature 3 Press until the word TIMER is displayed on the LCD the two left hand digits in the...

Page 11: ...bath is in the Standby mode 2 Press The actual bath temperature is displayed on the upper readout and the set point temperature is displayed on the lower readout Both values are displayed to one decim...

Page 12: ...ded to inform you about how much time has elapsed since the timer reached 00 00 To clear the Timer Value place the unit in Standby mode and press to return from Standby mode Using the Temperature Pres...

Page 13: ...T on the LCD will flash 1 Place the Mains Switch on the rear of the unit in the Off position 2 Press and hold while returning the Mains Switch to the On position 3 When the unit enters Standby the thr...

Page 14: ...event that electrical power is lost while your Water Bath is in use it will begin operating automatically at the set point temperature once power is restored The timer will begin counting up from the...

Page 15: ...sired set point temperature is displayed on lower readout Press or allow the display to timeout to accept C or F on lower readout Informational Indicates that the temperature unit is being changed Pre...

Page 16: ...tion your Water Bath incorporates redundant over temperature safeties These over temperature low liquid level safeties automatically remove power from the heater in the event that the bath temperature...

Page 17: ...uid as required Insufficient heating Improper line voltage Recent change in set point or heat load Bath cover not in place Check that line voltage meets specifications Allow sufficient time for bath t...

Page 18: ...liter 10 6 x 11 6 x 6 26 9 x 29 5 x 15 2 cm 1000 WB10 1 8 6 A WB10 2 4 5 A 20 liter 9 5 x 17 5 x 5 24 1 x 44 5 x 15 2 cm 1400 WB20 1 12 0 A WB20 2 6 0 A 28 liter 9 5 x 17 x 8 24 1 x 43 2 x 20 3 cm 14...

Page 19: ...distilled H2 0 5 C 0 92 3 85 5 to 90 C 10 to 107 C polycool MIX 25 50 50 mix with distilled H20 20 C 0 78 3 26 25 to 100 C 30 to 115 C polycool MIX 25 30 70 mix with distilled H20 0 C 0 89 3 72 0 to...

Page 20: ...have to be frequently replenished Caution must be taken to stay well below the fluid s flashpoint Application Notes For optimum results maintain fluid level throughout the operating period adding flui...

Page 21: ...ycling It is also your responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect the persons involved in the disposal and recyclin...

Page 22: ...t should be shipped transportation charges pre paid in either its original packaging or packaging affording an equal degree of protection to the indicated service center To insure prompt handling the...

Reviews: