10
110-031 PSC/EN
Normal Operation
Keypad Controls
Power
Turns the Controller On and Off.
Menu
Used to select and set temperature
units, timer time duration, high limit
temperature and calibration offset. Also
used to display the firmware version ID.
Timer
Used to start, pause and reset the
timer.
Selection
Used to increase/decrease temperature
set point and other operational settings/
values.
Temperature
Presets
(1, 2, 3)
Used to select one of three user-
defined set point temperatures and
optional time duration.
Turning Your Water Bath On
1. Place the Mains Switch on the rear of the unit in the On position. The
three decimal points in the upper readout light to indicate that the
Water bath is in the Standby mode.
2. Press . The actual bath temperature is displayed on the upper
readout and the set point temperature is displayed on the lower
readout. Both values are displayed to one decimal point. If the actual
bath temperature is lower than the set point temperature, the unit will
automatically begin heating and the heating symbol will light.
Adjusting the Set Point Temperature
This is the temperature at which the fluid in your bath will be maintained.
It may be set to one-tenth of a degree. The factory default set point is
+20.0ºC/+68ºF.
1. Press the or button. The word SET on the LCD will begin
flashing.
2. Use the and buttons to adjust the set point temperature to the
desired value. Press or allow the display to timeout (~10 seconds)
to accept the new value; the word SET on the LCD will be lit steady.
www.calcert.com
1.888.610.7664
0
5
10
15
20
25
30