110-392
16
Displaying and Adjusting the Ambient Tracking Offset
NOTE:
Ambient tracking is an optional function that may or may not be available on your Chiller. It permits
you to control fluid temperature based on room or machine temperature plus or minus a user-adjustable
offset temperature.
When the optional ambient tracking probe is installed and enabled (AtC), the ambient tracking offset rather than
the set point temperature is displayed when the Select/Set Knob on the front panel is pressed.
To change the displayed offset value, rotate the Select/Set Knob until the desired offset value is displayed. An
offset value from -5.0°C to +5.0°C (-9.0° to +9.0°F) may be entered. The setting is accepted after the Select/Set
Knob is pressed a second time or automatically after a few seconds of inactivity.
Displaying and Adjusting the Remote Control Temperature
NOTE:
Remote temperature control is an optional function that may or may not be available on your Chiller.
It permits you to control cooling based on the temperature of the external process.
When the optional remote control external probe is installed and enabled (rP), the external temperature set point
is displayed when the Select/Set Knob on the front panel is pressed.
To change the external temperature set point, press and then rotate the Select/Set Knob until the desired set
point temperature is displayed. The setting is accepted after the Select/Set Knob is pressed a second time or
automatically after a few seconds of inactivity.
Selecting the Pressure / Flow Rate Display and Units
The Chiller can set to display either fluid pressure (in PSI or kPa) or flow rate (in GPM or LPM). Pressing the
Units/Menu Select Button briefly toggles through the available selections.
NOTE:
Metric pressure reading output is displayed in kPA and must be multiplied by 100 for Pa.
NOTE:
The flow rate readout is intended as a reference only. If accurate flow readings are required, an
external flow meter is recommended.
Summary of Contents for DuraChill 58 Series
Page 1: ...User Guide DuraChill Air and Water Cooled 1 5 HP Chillers 110 392 27 June 2016...
Page 2: ......
Page 32: ...110 392 30 Electrical Specifications...
Page 39: ...110 392 37 Appendix Flow Schematic All Models...
Page 40: ...110 392 38 Electrical Diagram 208 230V 1PH 50 60Hz Models without heater option...
Page 41: ...110 392 39 Electrical Diagram 208 230V 1PH 50 60Hz Models with heater option...
Page 42: ...110 392 40 Electrical Diagram 208 230V and 380 460V 3PH 50 60Hz Models without heater option...
Page 43: ...110 392 41 Electrical Diagram 208 230V and 380 460V 3PH 50 60Hz Models with heater option...
Page 45: ...110 392 43 DI Control Wiring H2 H3 Option Wiring H2 24 VDC H3 12 VDC...