110-392
19
Menu Item
Description
Choices / Ranges
Default Setting
CC
Current Control
(optional) — Allows setting of the set point
temperature via a 4-20 mA analog signal.
yES — Enables current control.
NO — Disables current control.
Yes or No
No
PC
Communications Baud Rate
— Selects the baud rate for
serial (RS232/RS485) communication.
24 (2400), 48 (4800),
96 (9600) or 192 (19200)
9600
The current Fuse Bits (Fb) setting appears after the PC setting and can be viewed, but not changed. To change Fuse Bits
(Fb) operational parameters, electrical power must be turned OFF and a special key stroke combination entered. See
Fuse
Bits (Fb)
for detailed instructions.
Fb
Fuse Bits
Input
Logic
State
h00
h01
h02
h04
h05
h06
Remote ON / OFF
dry contact
Closed
Off
On
N/A
Off
On
N/A
Open
On
Off
On
Off
24 VDC Remote
Control Voltage
24 VDC
Off
Off
On
Off
N/A
On
0 VDC
On
On
Off
On
Off
Water Level Float
Switch
Open
OK
OK
OK
Low
Low
Low
Closed
Low
Low
Low
OK
OK
OK
Summary of Contents for DuraChill 58 Series
Page 1: ...User Guide DuraChill Air and Water Cooled 1 5 HP Chillers 110 392 27 June 2016...
Page 2: ......
Page 32: ...110 392 30 Electrical Specifications...
Page 39: ...110 392 37 Appendix Flow Schematic All Models...
Page 40: ...110 392 38 Electrical Diagram 208 230V 1PH 50 60Hz Models without heater option...
Page 41: ...110 392 39 Electrical Diagram 208 230V 1PH 50 60Hz Models with heater option...
Page 42: ...110 392 40 Electrical Diagram 208 230V and 380 460V 3PH 50 60Hz Models without heater option...
Page 43: ...110 392 41 Electrical Diagram 208 230V and 380 460V 3PH 50 60Hz Models with heater option...
Page 45: ...110 392 43 DI Control Wiring H2 H3 Option Wiring H2 24 VDC H3 12 VDC...