27
Attachment
А
Order sheet of modes operation and functions of the devices:
PM1703
М
-
О
, PM1703
М
-
О
1, PM1703
М
-
О
1A, PM1703
МА
, PM1703
МВ
,
PM1703GN
О
, PM1703GN
А
, PM1703GN
В
,
#
Modes and functions
On/off
----------
set value
Note
1
Search mode (gamma, indication "s
-1
")
Can be ON autonomous or combined with modes 2 or 3
2
Measurement mode
(indication µSv/h)
Can be ON autonomous or combined with modes 1 or 5
3
Measurement mode
(indication µR/h)
Can be ON autonomous or combined with modes 1 or 5
4
Gamma registration mode:
Modes of storage of counting pulse (gamma,
indications
-1
) and average count rate at storage time
Only for device PM1703
М
-
О
1
Can be ON additionally to modes 1, 2 or 3
5
Mode 0…..9:
Mode of relative indication
(0-9)
Can be ON autonomous or combined with modes 2 or 3
6
Auto calibration
Impossible in mode 5
7
Search in measurement mode
Possible only in combining of modes 1 and 2
Or 1 and 3
8
Installation of thresholds DER for mode 5
- Search threshold
- threshold DER (threshold of safety)
threshold: 15 – 7777
µ
R/h (0.15-77.77
µ
Sv/h)
advisable30
µ
R/h (0,3
µ
Sv/h)
threshold 10 – 7000
µ
R/h (0,1-70
µ
Sv/h)
advisable - 3mR/h (30
µ
Sv/h)
9
Installation of thresholds DER for modes 1-3
- threshold DER (threshold of safety))
threshold- 1 – 7000
µ
R/h (0.01 – 70
µ
Sv/h)
advisable- 3mR/h (30
µ
Sv/h)
10
Switch Audio/Vibr:
Fast switch of alarm mode:
audio-light or vibr-light
Возможность
быстрого
переключения
вида
сигнализации
без
входа
в
режим
установок
11
Changing of volume level:
Adjustment of level of audio alarm by buttons
Possibility of change of volume level of audio alarm in settings mode