Polimaster PM 1703GNA Operation Manual Download Page 24

 

24 

-time  and  indicator  readings  in  case  of  exceeding  the  operation  threshold  of  the 

gamma channel

-  time  and  indicator  readings  in  case  of  exceeding  the  operation  threshold  of  the 

neutron channel

;

 

In settings of indicator/program 

- to check and/or set up working parameters of the indicator:

 

- to switch on the sound/or vibrating indicators; 
- to synchronize the time and the date of the indicator with the current time and the PC date on the 

moment of information exchange – automatically during every communication of the indicator with the PC;  

- to set up the values of the values of the consecutive time intervals for storing the current values of 

MED in the power-independent indicator storage;  

- to change the password for entrance to the file of parameters (the initial password -1); 
- to check and set up the fixed threshold by MED, if exceeded, the indicator provides a light, sound 

and/or vibrating signals;  

-  to  check  the  preset  or  set  up  new  values  of  n-factors  by  each  channel  determining  the  operation 

thresholds (minimal values of gamma and neutron radiation detection);  

- automatic calibration should be switched on/off; 
 

 

TECHNICAL MAINTENANCE 

The technical maintenance of the indicator: 
-  performance  of  preventive  works  (external  inspection,  dust  removal  and  deactivation,  check  of 

indicator workability (see indicator on/off). Deactivation is performed by wiping with clothes moistened in 
ethyl  spirit.  

-replacement of a supply cell  in the explosion hazard area is strictly prohibited). 
In case of visible mechanical damages of the casing and protection glass of the LCD indicator (dents, 

burrs, cracks) operation of the indicator is prohibited.  

 
POSSIBLE MALFUNCTIONS 
 

Typical malfunctions 

Troubleshooting 

• 

The indicator can not be switched on 

Wrong installation of the power cell 

• 

The LDC indicates the mark        "    " 

The power cell is to be replaced 

 

Other malfunctions shall be eliminated by the manufacturer. 

 

 

STORAGE AND TRANSPORTATION 
Storage 

Indicators  must  be  stored  at  storehouses  in  the  manufacturer’s  package  at  the  ambient  temperature 

from minus 15 to plus 50 

°

С

 and relative humidity by 95 % at temperature 35 

°

С

. Storage duration shall not 

exceed the average service life of the indicator – 8 years. 

Indicators  without  packages  should  be  stored    at  the  ambient  temperature  from    10  to  35 

°

С

  and 

relative humidity 80 % at temperature 25

 

 

°

С

In the storage premises there shall be no dust, vapours of acids and alkalis, aggressive gases and other 

harmful admixtures causing corrosion. 

 
Transportation 

Indicators in the packaged form and switched off state allow transportation by any closed transport 

type at the ambient temperature from minus 50 to plus 50

о

C. 

The packed indicators must be fixed in a vehicle. Placement and fastening of the packed indicators in 

the  vehicle  shall  provide  their  stable  position,  exclude  possibility  of  shocks  on  each  other  ,  as  well  as  on 
vehicle walls.  

Summary of Contents for PM 1703GNA

Page 1: ...POLIMASTER PERSONAL RADIATION DETECTORS 1703 1703GN 1703 1703GN 1703 1703GN 1703 1 1703 1 OPERATION MANUAL http www polimaster com...

Page 2: ...Indication mode of neutrons 17 Indication mode of the average speed of and count of neutrons over the time of accumulation 17 Quick switching of sound or vibration indication 17 Communication mode by...

Page 3: ...erred to as MED The indicator is not a meter The indicator can be operated both in premises and in the open air The indicator can be used by a wide circle of consumers who are by their mode of activit...

Page 4: ...oth type Possibility of identification of the radio nuclide composition of the substance with the aid of ATTENTION The indicators may differ by the totality of included activated operation modes Opera...

Page 5: ...1 Personal radiation detector 1703GNO 1 Personal radiation detector 1703GNA 1 Personal radiation detector 1703GN 1 Disk software on CD 1 1 1 1 1 1 1 1 Panasonic POWER LINE LR6 power supply cell 1 1 1...

Page 6: ...neutron radiation registration 1703GN 1703GN 1703GN From heat to 14 0 MeV Range of indication of MED of photon radiation 1703 1 1703 1A 1703 1703GN 1703 1703 1703GN 1703GN 0 01 9999 mc v h 1 1000 mR h...

Page 7: ...corded events to the indicator storage by 1000 Terms of operation range of ambient temperatures relative humidity From minus 30 C to 50 22 F to 122 F LCD from minus 15 C to plus 50 by 95 at 35 C 95 F...

Page 8: ...quirements of IEC 60079 0 owing to the following designing and circuit technical solutions provision of the normal degree of mechanical strength of the indicator casing in accordance with the requirem...

Page 9: ...irected inside the indicator the cover of the supply section is to be restored When the supply cell is installed the indicator is automatically switched on During switching on and during operation of...

Page 10: ...ion test CAL OL OFF 1 3 etc an indication type sound or vibration 6 the analogue scale consisting of 19 segments is intended for time indication to completion of the internal processor tests reduction...

Page 11: ...t the LCD for several seconds After completion of the tests the indicator must be shifted to the calibration mode by the background level except for devices having the search mode 0 9 the analogue sca...

Page 12: ...ing analogue scale Tests are performed The time to completion of the test is displayed in relative units on the analogue scale in the form of the reducing number of the indicated segments After the te...

Page 13: ...function may be switched on in the mode of communication with the PC Automatic calibration makes it possible to automatically retain the high sensitivity at reduction of the background level and avoid...

Page 14: ...he threshold value indication light sound and or vibrating is switched on 3 The operation threshold for the neutron channel21 c The minimum detection level calculated in the calibration mode and takin...

Page 15: ...ization can be continued If necessary these actions cam be repeated several times up to location of II During localization of the source of neutron2 or mixed gamma and neutron radiations the sound and...

Page 16: ...ation source 2 The safety threshold the 2nd threshold by MED is only set in the mode of communication with a PC by the user administrator The range of threshold setting is equal to 10 7000 R h 0 1 70...

Page 17: ...E SPEED OF NEUTRON RADIATION INDICATION DURING ACCUMULATION 2 In this mode in the upper line of LCD the average count time is shown during accumulation time s 1 The lower LCD line shows the value of t...

Page 18: ...the indicator enters after approximately 1 minute the sleeping mode and being so LCD will have the following forms the radio channel is on no communication with the P the radio channel is on communica...

Page 19: ...LIGHT push button increase the preset value of n factor with pitch 0 1 Subsequent pressings of the MODE push button increase the preset value of n factor with pitch 0 1 If the push buttons are kept pr...

Page 20: ...of the searching threshold by MED will blink by indicating its possible change If the LIGHT button was not pressed during the said time interval the indicator automatically returns to the working mode...

Page 21: ...ode MODE OF INDICATION SETTINGS ON OFF OF SOUND AND OR VIBRATING INDICATION The state selection on off of vibrating and sound indicators from the front panel is possible if this mode is allowed at set...

Page 22: ...ce then at entering the mode of sound volume setting Aud note will be indicated on the LCD as well as the preset sound indicator value from 1 to 5 For changing the sound volume a LIGHT push button sho...

Page 23: ...omatic start up is supported Operation with the PP is described in th Help site in the attached textual document which are installed along with the user s software For switching on the communication m...

Page 24: ...tion is performed by wiping with clothes moistened in ethyl spirit replacement of a supply cell in the explosion hazard area is strictly prohibited In case of visible mechanical damages of the casing...

Page 25: ...anufacturer s permit The warranty does not concern the indicators at expiry of the warranty operation term if the indicator is purchased within the limits of the warranty storage period in case of the...

Page 26: ...nconformity for which Polimaster is responsible Purchaser shall inform Polimaster of the nature of such defect or nonconformity in reasonable detail and shall request authorization from Polimaster to...

Page 27: ...to modes 1 2 or 3 5 Mode 0 9 Mode of relative indication 0 9 Can be ON autonomous or combined with modes 2 or 3 6 Auto calibration Impossible in mode 5 7 Search in measurement mode Possible only in co...

Page 28: ...t threshold count neutrons Only for mode 5 threshold 1 0 99 0 s 1 with interval 0 5 advisable 1 5 s 1 Additional settings 14 Interval of history record min 15 Resolution of coefficient variation n 16...

Reviews: