Polimaster PM 1703GNA Operation Manual Download Page 13

 

13 

The processor counts the value of 

the gamma channel operation threshold

 

Operation threshold = N + n * 

N

 

N – average speed of counting impulses over the calibration time 

n

 – number of mean square deviations ( 

factor of the gamma channel). 

 

factor  changes  the  value  of 

the  operation  threshold 

(the  minimum  detection  level),  the  less   

factor,  the  less  threshold  value  is  and  the  higher  indicator  sensitivity  is.   

factor  is  set  by  the  user  in  the 

mode  of  settings,  if  this  is  permitted    in  the  communication  mode  with  PC  by  the  user-administrator  (the 
manufacturer sets the value of n factor equal to 5,3). The range of factor setting is equal from 1 to 9,9 with 
discreteness 0,1. 

 

For recalibration of the indicator by the background level the 

MODE

  button should be pressed (to 

be  kept  pressed  for  more  than  2  s),  until  “CAL”  message  is  indicated  on  the  LCD,  then  the  push  button 
should be released. The analogue scale will also be indicated on the LCD along with the number of segments 
increasing in time. 

 

The automatic calibration function may be switched on  in the mode of communication with the PC. 

Automatic  calibration  makes  it  possible  to  automatically  retain  the  high  sensitivity  at  reduction  of  the 
background level and avoid false operations with its slow increase.  

 

After  calibration  completion  the  indicator  is  automatically  shifted  to  the  operating  mode  in 

accordance with the chart of orders (see attachment 

А

).

 

 
SEARCHING MODE. 
DETECTION  AND  LOCALIZATION  OF  SOURCES  OF  GAMMA  AND  NEUTRON 

RADIATIONS. 

Being in the searching mode the indicator shows the speed of counts of the gamma channel in the LCD 

upper line, s

-1

 

The following is indicated in the lower line: 

РМ

 1703

М

-

О

РМ

 1703

М

-

О

1

А

 

РМ

 1703

М

-

О

 

РМ

 1703

МА

 

РМ

 1703

МВ

 

 

 

• 

the average  statistical  indication error  of  the average  count 
of gamma radiation in per cents;  

 

РМ

 1703GN

О

 

РМ

 1703GN

А

 

РМ

 1703GN

В

 

 

 

 

• 

the average speed of count of neutron radiation  in s

-1;

 

 

All modifications 

 

• 

If at indicator operation in the searching mode the average 
speed of count of the gamma channel exceeds the upper 
limit of indication, 

"OL" 

is indicated on the LCD

.

 

РМ

 1703GN

О

 

РМ

 1703GN

А

 

РМ

 1703GN

В

 

 

 

• 

If at indicator operation in the searching mode the average 
speed of count of the gamma channel exceeds the upper 
limit of indication, “999" blinking message is indicated on 
LCD. 

Summary of Contents for PM 1703GNA

Page 1: ...POLIMASTER PERSONAL RADIATION DETECTORS 1703 1703GN 1703 1703GN 1703 1703GN 1703 1 1703 1 OPERATION MANUAL http www polimaster com...

Page 2: ...Indication mode of neutrons 17 Indication mode of the average speed of and count of neutrons over the time of accumulation 17 Quick switching of sound or vibration indication 17 Communication mode by...

Page 3: ...erred to as MED The indicator is not a meter The indicator can be operated both in premises and in the open air The indicator can be used by a wide circle of consumers who are by their mode of activit...

Page 4: ...oth type Possibility of identification of the radio nuclide composition of the substance with the aid of ATTENTION The indicators may differ by the totality of included activated operation modes Opera...

Page 5: ...1 Personal radiation detector 1703GNO 1 Personal radiation detector 1703GNA 1 Personal radiation detector 1703GN 1 Disk software on CD 1 1 1 1 1 1 1 1 Panasonic POWER LINE LR6 power supply cell 1 1 1...

Page 6: ...neutron radiation registration 1703GN 1703GN 1703GN From heat to 14 0 MeV Range of indication of MED of photon radiation 1703 1 1703 1A 1703 1703GN 1703 1703 1703GN 1703GN 0 01 9999 mc v h 1 1000 mR h...

Page 7: ...corded events to the indicator storage by 1000 Terms of operation range of ambient temperatures relative humidity From minus 30 C to 50 22 F to 122 F LCD from minus 15 C to plus 50 by 95 at 35 C 95 F...

Page 8: ...quirements of IEC 60079 0 owing to the following designing and circuit technical solutions provision of the normal degree of mechanical strength of the indicator casing in accordance with the requirem...

Page 9: ...irected inside the indicator the cover of the supply section is to be restored When the supply cell is installed the indicator is automatically switched on During switching on and during operation of...

Page 10: ...ion test CAL OL OFF 1 3 etc an indication type sound or vibration 6 the analogue scale consisting of 19 segments is intended for time indication to completion of the internal processor tests reduction...

Page 11: ...t the LCD for several seconds After completion of the tests the indicator must be shifted to the calibration mode by the background level except for devices having the search mode 0 9 the analogue sca...

Page 12: ...ing analogue scale Tests are performed The time to completion of the test is displayed in relative units on the analogue scale in the form of the reducing number of the indicated segments After the te...

Page 13: ...function may be switched on in the mode of communication with the PC Automatic calibration makes it possible to automatically retain the high sensitivity at reduction of the background level and avoid...

Page 14: ...he threshold value indication light sound and or vibrating is switched on 3 The operation threshold for the neutron channel21 c The minimum detection level calculated in the calibration mode and takin...

Page 15: ...ization can be continued If necessary these actions cam be repeated several times up to location of II During localization of the source of neutron2 or mixed gamma and neutron radiations the sound and...

Page 16: ...ation source 2 The safety threshold the 2nd threshold by MED is only set in the mode of communication with a PC by the user administrator The range of threshold setting is equal to 10 7000 R h 0 1 70...

Page 17: ...E SPEED OF NEUTRON RADIATION INDICATION DURING ACCUMULATION 2 In this mode in the upper line of LCD the average count time is shown during accumulation time s 1 The lower LCD line shows the value of t...

Page 18: ...the indicator enters after approximately 1 minute the sleeping mode and being so LCD will have the following forms the radio channel is on no communication with the P the radio channel is on communica...

Page 19: ...LIGHT push button increase the preset value of n factor with pitch 0 1 Subsequent pressings of the MODE push button increase the preset value of n factor with pitch 0 1 If the push buttons are kept pr...

Page 20: ...of the searching threshold by MED will blink by indicating its possible change If the LIGHT button was not pressed during the said time interval the indicator automatically returns to the working mode...

Page 21: ...ode MODE OF INDICATION SETTINGS ON OFF OF SOUND AND OR VIBRATING INDICATION The state selection on off of vibrating and sound indicators from the front panel is possible if this mode is allowed at set...

Page 22: ...ce then at entering the mode of sound volume setting Aud note will be indicated on the LCD as well as the preset sound indicator value from 1 to 5 For changing the sound volume a LIGHT push button sho...

Page 23: ...omatic start up is supported Operation with the PP is described in th Help site in the attached textual document which are installed along with the user s software For switching on the communication m...

Page 24: ...tion is performed by wiping with clothes moistened in ethyl spirit replacement of a supply cell in the explosion hazard area is strictly prohibited In case of visible mechanical damages of the casing...

Page 25: ...anufacturer s permit The warranty does not concern the indicators at expiry of the warranty operation term if the indicator is purchased within the limits of the warranty storage period in case of the...

Page 26: ...nconformity for which Polimaster is responsible Purchaser shall inform Polimaster of the nature of such defect or nonconformity in reasonable detail and shall request authorization from Polimaster to...

Page 27: ...to modes 1 2 or 3 5 Mode 0 9 Mode of relative indication 0 9 Can be ON autonomous or combined with modes 2 or 3 6 Auto calibration Impossible in mode 5 7 Search in measurement mode Possible only in co...

Page 28: ...t threshold count neutrons Only for mode 5 threshold 1 0 99 0 s 1 with interval 0 5 advisable 1 5 s 1 Additional settings 14 Interval of history record min 15 Resolution of coefficient variation n 16...

Reviews: