the searching function is additionally included in the measuring mode 

(see the order chart – attachment 

A).

 

 

SEARCH MODE "0-9". 

In  the  searching  mode  "0-9"  digits  from  0  to  9  are  indicated  on  the  LCD  display.  Compliance  of 

MED values  by 

γ

 - channel and digits indicated on the indicator LCD is shown in the table. 

Indicated figure 

Range of 

indicated MEG, 

mc

З

v/h 

From 

background 

to  3,6 

3,6 

<

7,2 

7,2 

<

10,8 

10,8 

<

14,4 

14,4 

<

18,0 

18,0 

<

21,6 

21,6 

<

25,2 

25,2 

<

28,8 

28,8 

<

32,4 

32,4 

<

36,0 

The analogue scale is in parallel filled in. The number of illuminating segments of the analogue scale  

is proportional to the digit induced on the LCD. 
In  the  searching  mode  "0-9"  operation  of  the  indicator  is  carried  out  in  the  following 
thresholds:  

1)

 The searching 

threshold by MED 

 is set in the mode of settings from the front 

panel  of  the indicator, as  well  as  in  the  mode  of  communication  with  a PC  by  the  user-administrator.  The 
range  of  threshold  setting  is  equal  to    15  –  7777 

µ

R/h  (0.15-77.77

µ

Sv/h),  recommended  -  30

µ

R/h  (0,3 

µ

Sv/h)). The preset searching threshold exceeding by MED is accompanied with light and sound signaling. 

Mechanical vibrations  inside the device (casing trembling) are felt with the switched on vibrating indicator 
instead  of  sound  signals.  Frequency  of  signals  increases  by  device  approaching  to  the  gamma  radiation 
source.  

 

2)

 

The  safety  threshold  –  the  2

nd

  threshold  by  MED

,  is  only  set  in  the  mode  of  communication 

with  a  PC  by  the  user  administrator.  The  range  of  threshold  setting  is  equal  to  10  –  7000 

µ

R/h  (0.1-

70

µ

Sv/h),  the  recommended  one  –  3mR/h  (30 

µ

Sv/h)).  The  light,  sound  and/or  vibrating  signals  are 

provided in case of exceeding the safety threshold. Being so, single tone periodical signals are provided with 
the permanent interval and duration of 1 s what makes it possible for the user  to differentiate  the indication 
in case of exceeding the searching threshold by MED from the safety threshold.  

 

Compliance of speed values by neutron

2

1

 channel and digits indicated on the device LCD 

is shown in the following table. 

 

Indicated figure 

Count speed 

From 

backg
round  

to 2 

<

<

<

<

10 

10 

<

17 

17 

<

24 

24 

<

31 

31 

<

41 

41 

The fixed threshold of counting neutrons  for the searching mode "0-9" (the range of threshold setting 

is  equal  to    1.0  –  99.0  s

-1

,  recommended  –  1.5  s

-1

)  is  set  in  the  mode  of  settings  from  the  front  indicator 

panel, as well as in the communication mode with a PC by the user-administrator.  

 
GAMMA REGISTRATION MODE (only for 

РМ

1703

М

-

О

1) 

 

In this mode in the upper line of LCD a 

γ

 is shown and the quantity of accumulated count impulses is 

indicated in the form of: 

 

ХХХХ

 – in case of indication by the value of  9999 impulses  

 

Х

.

ХЕХ

 – in case of indication above  9999 impulses, wherein 

Х

 is any figure 

from  0 to 9, 

ЕХ

 is 10

х

Time of accumulation of impulses in hours is indicated in the lower LCD line. 

 
 
 
 

                                                           

2

 

РМ

1703GNO, 

РМ

1703GNA, 

РМ

1703GNB

 

Summary of Contents for PM 1703GNA

Page 1: ...POLIMASTER PERSONAL RADIATION DETECTORS 1703 1703GN 1703 1703GN 1703 1703GN 1703 1 1703 1 OPERATION MANUAL http www polimaster com...

Page 2: ...Indication mode of neutrons 17 Indication mode of the average speed of and count of neutrons over the time of accumulation 17 Quick switching of sound or vibration indication 17 Communication mode by...

Page 3: ...erred to as MED The indicator is not a meter The indicator can be operated both in premises and in the open air The indicator can be used by a wide circle of consumers who are by their mode of activit...

Page 4: ...oth type Possibility of identification of the radio nuclide composition of the substance with the aid of ATTENTION The indicators may differ by the totality of included activated operation modes Opera...

Page 5: ...1 Personal radiation detector 1703GNO 1 Personal radiation detector 1703GNA 1 Personal radiation detector 1703GN 1 Disk software on CD 1 1 1 1 1 1 1 1 Panasonic POWER LINE LR6 power supply cell 1 1 1...

Page 6: ...neutron radiation registration 1703GN 1703GN 1703GN From heat to 14 0 MeV Range of indication of MED of photon radiation 1703 1 1703 1A 1703 1703GN 1703 1703 1703GN 1703GN 0 01 9999 mc v h 1 1000 mR h...

Page 7: ...corded events to the indicator storage by 1000 Terms of operation range of ambient temperatures relative humidity From minus 30 C to 50 22 F to 122 F LCD from minus 15 C to plus 50 by 95 at 35 C 95 F...

Page 8: ...quirements of IEC 60079 0 owing to the following designing and circuit technical solutions provision of the normal degree of mechanical strength of the indicator casing in accordance with the requirem...

Page 9: ...irected inside the indicator the cover of the supply section is to be restored When the supply cell is installed the indicator is automatically switched on During switching on and during operation of...

Page 10: ...ion test CAL OL OFF 1 3 etc an indication type sound or vibration 6 the analogue scale consisting of 19 segments is intended for time indication to completion of the internal processor tests reduction...

Page 11: ...t the LCD for several seconds After completion of the tests the indicator must be shifted to the calibration mode by the background level except for devices having the search mode 0 9 the analogue sca...

Page 12: ...ing analogue scale Tests are performed The time to completion of the test is displayed in relative units on the analogue scale in the form of the reducing number of the indicated segments After the te...

Page 13: ...function may be switched on in the mode of communication with the PC Automatic calibration makes it possible to automatically retain the high sensitivity at reduction of the background level and avoid...

Page 14: ...he threshold value indication light sound and or vibrating is switched on 3 The operation threshold for the neutron channel21 c The minimum detection level calculated in the calibration mode and takin...

Page 15: ...ization can be continued If necessary these actions cam be repeated several times up to location of II During localization of the source of neutron2 or mixed gamma and neutron radiations the sound and...

Page 16: ...ation source 2 The safety threshold the 2nd threshold by MED is only set in the mode of communication with a PC by the user administrator The range of threshold setting is equal to 10 7000 R h 0 1 70...

Page 17: ...E SPEED OF NEUTRON RADIATION INDICATION DURING ACCUMULATION 2 In this mode in the upper line of LCD the average count time is shown during accumulation time s 1 The lower LCD line shows the value of t...

Page 18: ...the indicator enters after approximately 1 minute the sleeping mode and being so LCD will have the following forms the radio channel is on no communication with the P the radio channel is on communica...

Page 19: ...LIGHT push button increase the preset value of n factor with pitch 0 1 Subsequent pressings of the MODE push button increase the preset value of n factor with pitch 0 1 If the push buttons are kept pr...

Page 20: ...of the searching threshold by MED will blink by indicating its possible change If the LIGHT button was not pressed during the said time interval the indicator automatically returns to the working mode...

Page 21: ...ode MODE OF INDICATION SETTINGS ON OFF OF SOUND AND OR VIBRATING INDICATION The state selection on off of vibrating and sound indicators from the front panel is possible if this mode is allowed at set...

Page 22: ...ce then at entering the mode of sound volume setting Aud note will be indicated on the LCD as well as the preset sound indicator value from 1 to 5 For changing the sound volume a LIGHT push button sho...

Page 23: ...omatic start up is supported Operation with the PP is described in th Help site in the attached textual document which are installed along with the user s software For switching on the communication m...

Page 24: ...tion is performed by wiping with clothes moistened in ethyl spirit replacement of a supply cell in the explosion hazard area is strictly prohibited In case of visible mechanical damages of the casing...

Page 25: ...anufacturer s permit The warranty does not concern the indicators at expiry of the warranty operation term if the indicator is purchased within the limits of the warranty storage period in case of the...

Page 26: ...nconformity for which Polimaster is responsible Purchaser shall inform Polimaster of the nature of such defect or nonconformity in reasonable detail and shall request authorization from Polimaster to...

Page 27: ...to modes 1 2 or 3 5 Mode 0 9 Mode of relative indication 0 9 Can be ON autonomous or combined with modes 2 or 3 6 Auto calibration Impossible in mode 5 7 Search in measurement mode Possible only in co...

Page 28: ...t threshold count neutrons Only for mode 5 threshold 1 0 99 0 s 1 with interval 0 5 advisable 1 5 s 1 Additional settings 14 Interval of history record min 15 Resolution of coefficient variation n 16...

Reviews: