PNI Sensor Corporation
DOC#1014688 r09.2
TCM User Manual
Page 55
7.3.23
kPowerUpDone (frame ID 23
d
)
This frame confirms the TCM received a command to power up. The TCM will power up
when it receives any signal on the native UART Rx line. The frame has no payload. Since
the module was previously powered down which drives the RS-232 driver TX line low
(break signal), it is recommended to disregard the first byte.
7.3.24
kSetAcqParams (frame ID 24
d
)
This frame sets the sensor acquisition parameters in the TCM. The payload should contain
the following:
AcquisitionMode
This flag sets whether output will be presented in Continuous or Polled Acquisition
Mode. Its default value is 0 (FALSE), indicating Polled Mode. Polled Mode should
be selected when the host system will poll the TCM for each data set. Continuous
Mode (value 1 as TRUE) should be selected if the user will have the TCM output data
to the host system at a relatively fixed rate.
FlushFilter
Setting this flag to TRUE results in the FIR filter being flushed (cleared) after every
measurement. The default is FALSE.
Flushing the filter clears all tap values, thus purging old data. This can be useful if a
significant change in heading has occurred since the last reading, as the old heading
data would be in the filter. Once the taps are cleared, it is necessary to fully repopulate
the filter before data is output. For example, if 32 FIR taps is set, 32 new samples must
be taken before a reading will be output. The length of the delay before outputting data
is directly correlated to the number of FIR taps.
AcquireDelay
When operating in Continuous Acquisition Mode, the AcquireDelay sets the time
between samples taken by the module, in seconds. The default is 0.0 seconds, which
means the module will reacquire data immediately after the last acquisition. This is an
internal setting that is NOT tied to the time with which the module transmits data to the
host system. Generally speaking, the AcquireDelay is either set to 0, in which case the
TCM is constantly sampling, or set to equal the SampleDelay value. The advantage of
AcquisitionMode
UInt8
FlushFilter
UInt8
Payload
AquireDelay
SampleDelay
Float32
Float32