Installation, Operation and Maintenance Manual
Page 36
Medical Vacuum Plant
Table 8-8; PureLogic® Central Controller – Pump
Bar Graph
Status
Description
Idle pump
The pump is idle and ready to be called. The bar graph
is blank.
Lead pump
The central controller has assigned this pump to the
next one to run. This is indicated by a full line
underneath the blank bar graph.
Called pump
This pump is running. The bar graph is coloured.
Called pump, last one
called
This pump has last started to run. This is indicated by
a dotted line underneath the coloured bar graph.
Table 8-9; Central controller – Fault icons (more details in section Plant fault).
Icon
Status
Description
No valid pump
A pump controller is expected at this CAN
address.
No communication
No reply from the connected pump controller
within a pre-defined time.
Blinking network icon
The pump is running in Emergency Forced Local
(EFL).
No answer
Connected pump controller is not responding
correctly to the commands (e.g. no reaction on a run
command).
Not available
The pump is stopped and is counting out the
Minimum Stop Time to prevent too many motor
starts per hour. During this time the pump is not
Summary of Contents for HTM02-01
Page 1: ...Medical Vacuum Plant HTM2022 HTM02 01 Plant Systems Operation and Maintenance Manual ...
Page 2: ......
Page 73: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 55 ...
Page 74: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 18 ...
Page 75: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 55 ...
Page 76: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 20 ...
Page 77: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 55 ...
Page 78: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 22 ...
Page 80: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 24 ...
Page 81: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 55 ...
Page 82: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 26 ...
Page 83: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 55 ...
Page 84: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 28 ...
Page 85: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 55 ...
Page 87: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 55 ...
Page 88: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 32 ...
Page 89: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 55 ...
Page 90: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 34 ...
Page 91: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 55 ...
Page 92: ...Installation Operation and Maintenance Manual Medical Vacuum Plant Page 36 ...