5. Select the green check mark button.
NOTE
Contact your service technician for help if error message E332 (Severe
arcing across X-ray tube) reoccurs after a successful seasoning process.
13.1.5 Exposure Values (1600)
Steps
1. Go to 1600 Exposure Values to check the default kV and mA settings for
3D or Panoramic imaging.
The values can be adjusted as follows:
•
5 kV increments for 3D
•
2 kV increments for Panoramic
NOTE
The mA adjustment increments are measured according to a decimal
scale with divisions proportioned to the relevant mA measures.
13 Settings
88 Planmeca Viso
User's manual