11.12 Taking 3D face photo
Before you begin
Select the ProFace program from the main page by selecting Special
Programs > ProFace. For more information, see section "3D dental
programs" on page 50.
Make sure that you have selected the correct patient in the Planmeca
Romexis program.
Steps
1. Position the patient as described in section "Patient positioning" on page
54.
Adjust the patient position and the settings on the screen until the
Frankfort line and the patient's eye are correctly positioned as shown
in the image below.
NOTE
You can move the Frankfort line and eye markers using a finger (touch
screen) or the mouse cursor (virtual control panel).
2. Select the forward button in the bottom right corner.
Green lights flash on the control panel and exposure button when the
X-ray system is getting ready for an exposure.
The following screen displays.
The green lights stop flashing and stay on continuously when the X-ray
system is ready for an exposure. The following screen displays.
11 3D exposure
72 Planmeca Viso
User's manual