3D ENT programs, Planmeca Viso G7
Program
Preset volume
size
Volume range
Note
Nose
Ø80 × 80 mm
Ø70 × 50 mm – Ø120 × 100 mm
10 mm steps
Sinus
Ø130 × 130 mm
Ø100 × 100 mm – Ø160 × 160 mm
10 - 20 mm steps
Middle Ear
Ø50 × 50 mm
Ø30 × 30 mm – Ø60 × 60 mm
5 mm steps
75 µm voxel possible
only with volumes
smaller than 70 x 70 mm
Temporal
bone
Ø80 × 80 mm
Ø70 × 50 mm – Ø120 × 100 mm
10 mm steps
Airways
Ø90 × 100 mm
Ø80 × 80 mm – Ø100 × 140 mm
10 mm steps
Vertebrae
Ø80 × 100 mm
Ø80 × 80 mm – Ø100 × 140 mm
10 mm steps
3D programs, Planmeca Viso G5
Program
Preset volume
size
Volume range
Note
Tooth
Ø50 × 50 mm
Ø30 × 30 mm – Ø60 × 60 mm
5 mm steps
75 µm voxel possible
only with volumes
smaller than 70 x 70 mm
TMJ
Ø50 × 50 mm
Ø30 × 30 mm – Ø60 × 60 mm
5 mm steps
Teeth
Ø90 × 90 mm
Ø70 × 30 mm – Ø90 × 90 mm
10 mm steps
Jaw
Ø140 × 100 mm
Ø100 × 30 mm – Ø200 × 100 mm
10 mm steps
Face
Ø160 × 160 mm
Ø140 × 130 mm – Ø200 × 170 mm
10 mm steps
3D ENT programs, Planmeca Viso G5
Program
Preset volume
size
Volume range
Note
Nose
Ø80 × 80 mm
Ø70 × 50 mm – Ø90 × 90 mm
10 mm steps
Sinus
Ø130 × 130 mm
Ø100 × 100 mm – Ø200 × 150 mm
10 - 20 mm steps
Middle Ear
Ø50 × 50 mm
Ø30 × 30 mm – Ø60 × 60 mm
5 mm steps
75 µm voxel possible
only with volumes
smaller than 70 x 70 mm
Temporal
bone
Ø80 × 80 mm
Ø70 × 50 mm – Ø90 × 90 mm
10 mm steps
11 3D exposure
52 Planmeca Viso
User's manual
Summary of Contents for Viso
Page 1: ...Planmeca Viso user s manual EN 30005931 30005931...
Page 102: ...Patient supports 16 Cleaning and disinfection 98 Planmeca Viso User s manual...
Page 110: ......
Page 111: ......