Prostyle Intra X-ray Unit 27
TECHNICAL SPECIFICATIONS
User's manual
15
TECHNICAL SPECIFICATIONS
Generator
Constant potential, microprocessor
controlled,
operating frequency 66 kHz
X-ray tube
Toshiba D-0711SB
Focal spot size
0.7 mm
according to IEC 336
Cone diameter
60 mm (2.36 in.)
Rectangular 35 x 45 mm (1.38 x 1.77 in.)
Max. symmetrical radiation field
ø 60 mm at SSD 200 mm
ø 60 mm at SSD 300 mm
according to IEC 806
Total filtration
min. 2 mm Al equivalent at 70 kV
according to IEC 522
Inherent filtration
1 mm Al equivalent at 70 kV
according to IEC 522
Anode voltage
50, 52, 55, 57, 60, 63, 66, 70 kV, ±2 kV
Anode current
8 mA
Target angle
16°
Exposure times
0.01-3.20 sec., 23 steps
Reference current time product
8 mAs at 70 kV, 8 mA, 1 sec.
Lowest current time product
0.08 mAs at 8 mA, 0.01 sec.
Max. nominal anode voltage
70 kV
Max. electrical output
560 W at 70 kV, 8 mA
Electrical output at 0.1 sec.
560 W at 70 kV, 8 mA
Max. loading energy
1800 mAs/h at 70 kV
SSD (Source-Skin Distance) Standard/Long
Long with rectangular collimator
200 mm (8 in.)/300 mm (12 in.)
306 mm (12.04 in.)
Mains voltage
100 V~/110-115 V~/220-240 V~ ±10%
Mains frequency
50/60 Hz ±10%
Fusing
8 AT at 220-240V~
15 AT at 100 V~/110-115 V~
Duty cycle
1:15, automatic control
Electrical classification
Class I, type B
Weight
total 23 kg (51 lbs)
tube head 4.5 kg (10 lbs)
Ambient temperature
ope5°C - +40°C
storage -10°C - +50°C
Humidity
25% - 75%