9 INSTALLING THE DIGITAL SYSTEM
32 Planmeca ProOne
Installation Manual
9 INSTALLING THE DIGITAL SYSTEM
9.1 Connecting the Ethernet cable
The ProOne X-ray unit is connected to the local (10/100 Base) network and used from a
PC with the Romexis imaging software. Use an Ethernet cable to connect the X-ray unit to
the network.
Connect the Ethernet cable to the terminal on the back of the moving column. To do this,
remove the cover that protects the terminals if it is not already off. Remove the protective
cover by unscrewing the two screws at the bottom and then unlatching the cover at the
top.
First attach a ferrite bead (included in the delivery) to the Ethernet cable. Then connect the
cable to the right hand side terminal (marked ETH).
NOTE
Only the Ethernet cable can be connected to the right hand side terminal. Do not connect
any other cables to this terminal.
NOTE
The left hand side terminal is a USB port and it is reserved for a USB memory stick. Do not
plug any other USB devices into the USB port.
NOTE
Reattach the cover that protects the terminals by first latching the cover in place at the top
and then fastening the two screws at the bottom.
Back of
moving column
Ethernet cable
Screws that hold the protective cover in position
Ferrite bead
Summary of Contents for ProOne
Page 1: ...PlanmecaProOne EN 10016257_12 installation manual ...
Page 2: ......
Page 38: ...10 POST INSTALLATION CHECK LIST 36 Planmeca ProOne Installation Manual ...
Page 39: ...11 DIAGRAMS Installation Manual Planmeca ProOne 37 11 DIAGRAMS ...
Page 40: ......
Page 41: ......
Page 43: ......
Page 45: ......
Page 47: ......
Page 49: ......
Page 51: ......
Page 53: ......
Page 55: ......
Page 56: ......