8 INSTALLING THE EXPOSURE SWITCH
Installation Manual
Planmeca ProOne 29
NOTE
Only the exposure switch cable can be connected to the terminals in the middle. Do not
connect any other cables to these terminals.
NOTE
The left hand side terminal is a USB port and it is reserved for a USB memory stick. Do not
plug any other USB devices into the USB port.
Connect the other cable end to the exposure switch.
Back of
moving column
Exposure
switch cable
Screws that hold the protective cover in position
C
Ferrite bead
Exp_sw_wall5.eps
D
Summary of Contents for ProOne
Page 1: ...PlanmecaProOne EN 10016257_12 installation manual ...
Page 2: ......
Page 38: ...10 POST INSTALLATION CHECK LIST 36 Planmeca ProOne Installation Manual ...
Page 39: ...11 DIAGRAMS Installation Manual Planmeca ProOne 37 11 DIAGRAMS ...
Page 40: ......
Page 41: ......
Page 43: ......
Page 45: ......
Page 47: ......
Page 49: ......
Page 51: ......
Page 53: ......
Page 55: ......
Page 56: ......