Planmeca Proline EC Panoramic X-ray 15
PREPARATIONS FOR THE EXPOSURE
User's manual
Slide the cassette into the cassette carriage. Make
sure that you insert the cassette in the direction of the
arrow marked on the door and that the cassette is
pushed completely into the cassette carriage.
7.3 Preparing the patient
Ask the patient to remove any spectacles, hearing
aids, dentures, and personal jewellery such as ear-
rings, necklaces, and hairpins.
Place a protective lead apron over the patient’s back if
required.
Demonstrating the unit without taking an exposure
If the patient is nervous you may have to reassure
them by demonstrating the unit. This “dummy run”
should be carried out with the radiation switched off.
Press the Control key to activate the additional func-
tion mode. The indicator light will come on.
Press any of these keys to switch the radiation off. The
kV and mA values will clear from the displays which
indicates that the radiation has been switched off.
15 x 30 cm
THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT
AND OPERATOR UNLESS SAFE EXPOSURE FACTORS
AND OPERATING INSTRUCTIONS ARE OBSERVED
ATTENTION CET APPAREIL A RAYONS X PEUT ETRE
DANGEREUX POUR LE PATIENT ET L'OPERATEUR
SI LES FACTEURS DE SECURITE ET LE MODE
D'EMPLOI NE SONT PAS SCRUPULEUSEMENT
RESPECTES
WARNING
Proscan
P
L
A
N
M
E
C
A
1
5
x
3
0
c
m
THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT
AND OPERATOR UNLESS SAFE EXPOSURE FACTORS
AND OPERATING INSTRUCTIONS ARE OBSERVED
ATTENTION CET APPAREIL A RAYONS X PEUT ETRE
DANGEREUX POUR LE PATIENT ET L'OPERATEUR
SI LES FACTEURS DE SECURITE ET LE MODE
D'EMPLOI NE SONT PAS SCRUPULEUSEMENT
RESPECTES
WARNING
Proscan
P
L
A
N
M
E
C
A
TION 2,5 mm Al
ATION 14 mm Al
OTALE 2,5 mm equ. Al
TOTALE 14 mm equ. Al
Cassette carriage
Cassette
CTL
OR
kV
mA