PLANMECA INTRA X-RAY UNIT
2
/8 Planmeca Intra Mobile base
Installation instructions
2.
Place the column e.g. onto the mobile base package so that the generator back plate is horizontal (1).
Attach the wheel assemblies to the base plate with the M10x40 DIN 912 screws using the 8mm Allen
key (2). Place the base plate onto the package as shown on the figure below. Rotate the column holder
(located in the column) so that the base plate can be attached to the column (base plate’s and column’s
screw holes in line) (3). Make sure that the gap between the column edge and column holder is approx.
3mm (0.12 in.) so that the column is firmly attached to the base plate. Attach the base plate to the col-
umn with four M10x40 DIN 912 screws (4) using the 8mm Allen key. Lift the mobile base up.
mobile_intr
a2.eps
1
3
2
4
2
3 mm
Column
Column holder
Wheel assembly
Base plate
Mobile base package
Generator back plate
Make sure that the gap between the column edge and column holder is
approx. 3mm (0.12 in.) so that the column is firmly attached to the base plate.