38 -
STR561 - User manual
15 Sampling frequency
Sampling frequency of analogue / digital converter.
NB
: Increasing the conversion speed will slow down reading stability
(ex: for fast transients like pressure, it is advisable to increase sampling
frequency)
1200 Hz
0.83ms (Maximum speed conversion)
600 Hz
1.67ms
240 Hz
4.16ms
75 Hz
13.3ms
37.5 Hz
26.7ms
30.0 Hz
33.3ms
15.0 Hz
66.7ms
12.5 Hz (
Default
) (80ms Ideal for filtering noises 50 / 60 Hz)
2.5 Hz
400ms
1.86 Hz
533ms
1.18 Hz
851ms (Minimum speed conversion)
13.2 Weight management
Parameters to manage the device as a scale.
133 Max zero set
Set the max. weight value which is possible to reset by “>0<”, digital inputs
or serial.
0..Full scale (
Default
: 1000)
134 Autozero start
Sets the max. weight value which is possible to reset at starting by the
autozero function. If at starting the value of the revealed weight is less
than this value, the weight is automatically reset. To disable this function
keep the parameter to 0.
0..20% Full scale (
Default
: 0)