I2.2.1 IN-FIELD CALIBRATION PROCEDURE
OPERATION
Display
1
NONE
K900 in normal mode, not in counting mode.
2
LONG CAL KEY KEYING
K900 enters calibration mode, shows «CAL” and displays the calibration factor in use
instead of total. The words “Fact” and “USER” indicate which of the two factors (factory
or user) is currently in use
3
LONG RESET KEY KEYING
K900 shows “CAL” and the partial at zero. The meter is ready to perform in-field cali-
bration.
4
DISPENSING INTO SAMPLE CONTAINER
Without pressing any button,
start dispensing into the sample container.
Dispensing can be interrupted and started again at will. Continue dispensing until the
level of the fluid in the sample container has reached the graduated area. There is no
need to reach a preset quantity.
Indicated value Real value
5
5
SHORT RESET KEY KEYING
K900 is informed that the calibration dispensing operation is finished.
Make sure dispensing is correctly finished before performing this operation.
To calibrate K900, the value indicated by the partial totaliser (example 155.20) must be
forced to the real value marked on the graduated sample container. In the bottom left
part of the display an arrow appears (upwards and downwards), that shows the direc-
tion (increase or decrease) of the USER K FACTOR value change when the operations
6 or 7 are performed.
6
SHORT RESET KEY KEYING
Changes the direction of the arrow. The operation can be repeated as many times as
you wish
7
SHORT/LONG CAL KEY KEYING
The indicated value changes in the direction indicated by the arrow
- one unit for every short CAL key keying
- continually if the CAL key is kept pressed.
(for the first 5 units slowly and then quickly).
If the desired value is exceeded, repeat the operations from point (6).