23
Vertical System
Vertical Sensitivity
2mV/div -10V/div
Channel Voltage Offset
Range
2mV ~ 200mV: ±1.6V
206mV ~ 10V: ±40V
Vertical Resolution
8 bit
Channels 2
Analog Bandwidth
100MHz
BW Flatness at BNC
input
DC -10% of rated BW: +/- 1dB
10% - 50% of rated BW: +/- 2dB
50% - 100% of rated BW: + 2dB/-3dB
Lower Frequency Limit
(AC -3dB)
≤
10Hz(at input BNC)
Noise
Pk-Pk for 3K record
≤
0.6 Div for average of 10 Pk-Pk
readings, Fixed gain settings
≤
0.7 Div for average of 10 Pk-Pk
readings, Variable gain settings
SFDR including
harmonics (measured
with FFT)
>=35dB
DC Gain Accuracy
<
±3.0%: 5mv/div to 10V/div in Fixed
Gain Ranges
<
±4.0%: 2mv/div Variable Gain
Ranges
DC Measurement
Accuracy: All Gain
settings
≤
100mv/div
± [3%*
(
|r|offset|
)
+1% *
|offset| +0.2div+2mv]
DC Measurement
Accuracy:
All Gain settings
>100mv/div
± [3%*
(
|r|offset|
)
+1% *