20
S:
“Print Key” option set to “Save Picture”.
○
5
Back USB Device:
“
Back USB” supports two types of interface: USBTMC and Printer.
○
6
Trigger Setting:
Trigger Level
. Display the position of the current trigger
level, for
example:
;.
Trigger Type
. Display the current trigger type and trigger
condition.
Different trigger types have different marks, for example: means
triggered on Slop side in edge trigger.
○
7
Frequency Counter:
Display the firmware frequency of current waveform. To display it,
you should turn on the “Counter” in menu of “UTILITY”.
○
8
Trigger Position:
Use
HORIZONTAL POSITION Knob
to modify the parameter. Turn
clockwise or counterclockwise to make the red arrowhead move
right or left, which will respectively cause the decrease and
increase of the parameter in the message box at the lower-left
corner of the screen. Press down the knob to automatically reset
the parameter to zero as well as make the red arrowhead return to
its initial position.
○
9
Horizontal Time Base:
Represent the time of each grid on the horizontal axis of the screen.
You could revolve
HORIZONTAL SCALE Knob
to modify the
parameter which is variable from 2.5nS to 50S.
○
10
BW Limit:
If the current “BW Limit” is “On”, then the mark B displays at the
lower-corner of the screen, or nothing displays. When the vertical
scale is 2mV/div, the “BW Limit” turns on automatically.