Pichler WPD-A3 Manual Download Page 8

PICHLER Kunststofftechnik GmbH

Reports on the Display

Please get a note for the reports on the Display. If the reports not leave within 24 hours,
the fault must be clean.

Please read further faults possibilities.

Fehlerbehebung

Fault / malifunction

to absorb the fault

PH, or redox value is over-
or underline.

- check if the container with liquid full or empty

(Please refill or replace the liquid at time. !)

The flow gauge is marked
with „X“

- No flow, please check if the cone is down.
- The filter sieve in the measure cell is dirty, please clean it.

Best for use take a old toothbrush and clean under cold
water tap.

- The flow sensor is broken, please renew it.

It shows on the screen the
countdown.

- Sombody had a bath before.
- The circulation pump stopped more than 10 seconds.
- Short time before you made a calibration and the water tap

was closed short time before.

One of the status color
LED is lighting
yellow or red.

- Short time before you calibrate a eletrode, so the electrode

life time ended. Please replace a new electrode.

- In the mainscreen the gauge is changing, as example

only 30 days. This is a 30 days countdown, and within this
time must be replace the electrode.

- You had calibrate wrong, please repeat it.

Please take sure during the calibration the electrode was into
the reference cotainer with liquid.

The status LED are
lighting green.

- Everything is ok.

Options, Updates, Service

Optionen sind:      free clorine cell ( for commercial using ), µS-electrode

Softwareupdates: Only the factory or dealer can update the software.

Service:                If you have any questions please write an email to: [email protected]

page 7 to 7

2017/09

Summary of Contents for WPD-A3

Page 1: ...ng System WPD A3 PICHLER Kunststofftechnik GmbH Lauterbachstra e 19 84307 Eggenfelden phone telefax email Internet 0 87 21 96 90 10 0 87 21 96 90 20 office pichler de www pichler dosiertechnik de 2017...

Page 2: ...2 print setting save load factory setting contrast 3 further adjustments time dosing alert start delay 4 hand dosing manuel dosing 5 electrodes calibration restore Reset 5 6 report on screen 6 troubl...

Page 3: ...creen will be shown as a arrow symbole on the side Status LED for ORP Redox status LED for pH Status SMD LED RGB screen The SMD LED lamps provide you quick view status report operating report If a mea...

Page 4: ...the sign ON for saving the changings push again the OK button Bild 1 Note Please save every time if you change some settings Push the button load settings with Adjustment load ON OFF the OK button tha...

Page 5: ...30 min If you have a malifunction in your house current or you connect it on off will start at first the start delay and you must wait for dosing is active this setting time Same is when the circulat...

Page 6: ...ctrode calibration Close both water tap so can avoid water come out of the measure cell Screw out the electrodes pH and Redox ORP And put the electrodes for 5 10 minute into the calibration container...

Page 7: ...ry setting Select on menu reset picture 16 Choose the reset and push the button OK Now select between pH and redox electrode for reset and press the ok button again pic 16 page 6 to 7 2017 09 calibrat...

Page 8: ...before The circulation pump stopped more than 10 seconds Short time before you made a calibration and the water tap was closed short time before One of the status color LED is lighting yellow or red...

Reviews: