Pichler WPD-A3 Manual Download Page 4

settings

menu customer

menu service

back

step:   1

21.09.2013   SA

13:07

=

=

7.2
755

PH
Redox

main pict.

print out the value

If a printer installed, can print out the measure values.

the values:

-quick print
-every hours aoutmatic print
-can adjust 8 different times for automatic printing.

This is a option, and can work only with a seperate printer

Note!

via bluetooth connection.

settings

menu service

back

step:   3

menu customer

page 3 to 7

PICHLER Kunststofftechnik GmbH

Settings

Please choose the menu „menu customer“, Settings (

).

pic 1

Push the button „save settings“ with

Adjustment. saving: ON/OFF

the „OK“ button, than with the arrow up                          it shows
the sign „ON“, for saving the changings push again the „OK“ button.
(

)

Bild 1

Note: Please save every time if you change some settings.

Push the button „load settings“ with

Adjustment load:     ON/OFF

the „OK“ button, than push the arrow up                    it shows
the sign „ON“ for load the changings push again the „OK“ - button.

(this can restore to customer menu)

factory settings:     ON/OFF

It will reset to the factory setting. (but only the customer menu).

(The range of contrast can be adjust

display contrast:     1-150

1-150  Please choose the optimal step for you.

pic  1

settings

factory settings

save settings

An

load settings

display contrast         55

Aus

Aus

customer menu

settings

more settings

manuell dosing

back

calibrate

print                           off

2017/09

V4.03

V4.03

menu customer

back

Notification: Newest is a WLAN module for web view and printing. Please ask for the model our service.

Summary of Contents for WPD-A3

Page 1: ...ng System WPD A3 PICHLER Kunststofftechnik GmbH Lauterbachstra e 19 84307 Eggenfelden phone telefax email Internet 0 87 21 96 90 10 0 87 21 96 90 20 office pichler de www pichler dosiertechnik de 2017...

Page 2: ...2 print setting save load factory setting contrast 3 further adjustments time dosing alert start delay 4 hand dosing manuel dosing 5 electrodes calibration restore Reset 5 6 report on screen 6 troubl...

Page 3: ...creen will be shown as a arrow symbole on the side Status LED for ORP Redox status LED for pH Status SMD LED RGB screen The SMD LED lamps provide you quick view status report operating report If a mea...

Page 4: ...the sign ON for saving the changings push again the OK button Bild 1 Note Please save every time if you change some settings Push the button load settings with Adjustment load ON OFF the OK button tha...

Page 5: ...30 min If you have a malifunction in your house current or you connect it on off will start at first the start delay and you must wait for dosing is active this setting time Same is when the circulat...

Page 6: ...ctrode calibration Close both water tap so can avoid water come out of the measure cell Screw out the electrodes pH and Redox ORP And put the electrodes for 5 10 minute into the calibration container...

Page 7: ...ry setting Select on menu reset picture 16 Choose the reset and push the button OK Now select between pH and redox electrode for reset and press the ok button again pic 16 page 6 to 7 2017 09 calibrat...

Page 8: ...before The circulation pump stopped more than 10 seconds Short time before you made a calibration and the water tap was closed short time before One of the status color LED is lighting yellow or red...

Reviews: