5 Installation
U651 Rotation Stage
MP136E
Version: 1.3.0
17
5.2
Connecting the U651 to the Protective Earth Conductor
INFORMATION
Pay attention to the applicable standards for connecting the protective earth conductor.
On the U-651, there is an M4 hole next to the cable exit for connecting the protective earth
conductor. In the following figure, this hole is marked with the symbol for the protective earth
conductor .
Figure 4: M4 threaded hole for connecting the protective earth conductor
Figure 5: Connecting the protective earth conductor (profile view)
1
Base body of the U-651
2
Flat washer
3
Lock washer
4
Screw
5
Cable lug
6
Protective earth conductor
Requirements
You have read and understood the general notes on installation (p. 15).
The U-651 is
not