Technical Data 

Specifications 

 

P-725.1CDE2  P-725.1CL 

P-725.1CD 
P-725.1CA 

P-725.2CL 
P-725.2CD 
P-725.2CA 

P-725.4CL 
P-725.4CD 
P-725.4CA 

P-725.x0L 
open-loop 

version 

Unit 

Tolerance 

Active axes 

 

 

Motion and positioning

 

 

 

 

 

 

 

 

Integrated sensor 

Capacitive 

Capacitive 

Capacitive 

Capacitive 

 

 

Travel range at -20 to +120 V, open 

loop 

120 

150 

330 

460 

as P-725.xCL 

µm 

+20 % /  

-0 % 

Travel range, closed loop 

100 

100 

250 

400 

µm 

 

Resolution, open loop 

0.1 

0.3 

0.4 

0.5 

as P-725.xCL 

Nm 

Typ. 

Resolution, closed loop 

0.2 

0.65 

0.75 

1.25 

nm 

Typ. 

Linearity, closed loop 

0.01 

0.03 

0.03 

0.03 

Typ. 

Repeatability 

±5 

±5 

±5 

±5 

nm 

Typ. 

Tilt θ

X

 

10 

10 

as P-725.xCL 

µrad 

Typ. 

Tilt θ

Y

 

10 

20 

45 

45 

as P-725.xCL 

µrad 

Typ. 

Crosstalk in X 

20 

20 

20 

60 

as P-725.xCL 

Nm 

Typ. 

Crosstalk in Y 

50 

20 

40 

60 

as P-725.xCL 

Nm 

Typ. 

Mechanical properties

 

 

 

 

 

 

 

 

Stiffness in motion direction 

0.50 

0.23 

0.17 

0.12 

as P-725.xCL 

N/µm 

±20 % 

Resonant frequency, unloaded 

680 

470 

330 

230 

as P-725.xCL 

Hz 

±20 % 

Resonant frequency, under load, 150 

290 

185 

140 

120 

as P-725.xCL 

Hz 

±20 % 

Push/pull force capacity in positioning 

direction 

100 / 20 

100 / 20 

100 / 20 

100 / 20 

as P-725.xCL 

Max. 

Drive properties

 

 

 

 

 

 

 

 

Ceramic type 

PICMA® 

PICMA® 

PICMA® 

PICMA® 

as P-725.xCL 

 

 

Electrical capacitance 

3.2 

4.2 

6.2 

6.2 

as P-725.xCL 

µF 

±20 % 

Miscellaneous

 

 

 

 

 

 

 

 

Operating temperature range 

-20 to 80 

-20 to 80 

-20 to 80 

-20 to 80 

-20 to 80 

°C 

 

Material 

Stainless steel, 

aluminum 

Aluminum 

Aluminum 

Aluminum 

Aluminum 

 

 

Objective diameter 

39 

39 

39 

39 

39 

mm 

Max. 

Assembly 

Ring clamp 

QuickLock 

QuickLock 

QuickLock 

QuickLock 

 

 

Mass 

0.28 

0.215 

0.23 

0.23 

as P-725.xCL 

kg 

±5 % 

Sensor/voltage connector 

D-sub 7W2 

(m) 

CL version: 

LEMO 
Other: 
D-sub 7W2 

(m) 

CL version: 

LEMO 
Other: 
D-sub 7W2 

(m) 

CL version: 

LEMO 
Other: 
D-sub 7W2 

(m) 

LEMO 

(no sensor) 

 

 

Cable length 

1.5 

1.5 

1.5 

1.5 

1.5 

+50 mm /  

-0 mm 

Recommended electronics 

E-505, E-610, 

E-621, E-625, 

E-665, E-709, 

E-754 

E-505, E-610, 

E-621, E-625, 

E-665, E-709, 

E-754 

E-505, E-610, 

E-621, E-625, 

E-665, E-709, 

E-754 

E-505, E-610, 

E-621, E-625, 

E-665, E-709, 

E-754 

E-505, E-610, 

E-621, E-625, 

E-665, E-709, 

E-754 

 

 

The resolution of the system is limited only by the noise of the amplifier and the measuring technology because PI piezo nanopositioning systems are 

free of friction. 
All specifications based on room temperature (22 °C ±3 °C). 

Summary of Contents for P-725.1CDE2

Page 1: ...uf der Roemerstrasse 1 76228 Karlsruhe Germany Phone 49 721 4846 0 Fax 49 721 4846 1019 Email info pi de www pi ws P 725 P 725 PIFOC Long Travel Objective Scanner P 725 1CDE2 PIFOC Piezo Nanofocusing...

Page 2: ...er Cable with D sub Mechanics to LEMO Electronics 12 Adapter Cable with LEMO Mechanics to D sub Electronics 13 Unpacking 13 Installation 13 Connecting the P 725 to the Protective Earth Conductor 13 Mo...

Page 3: ...846 0 Fax 49 721 4846 1019 Email info pi de www pi ws Customer Service Department 30 Technical Data 31 Specifications 31 Maximum Ratings 33 Ambient Conditions and Classifications 33 Dimensions 34 Torq...

Page 4: ...of control technology drive technologies and suitable safety measures is assumed Typographic Conventions Symbol Label Meaning 1 2 Action consisting of several steps with strict sequential order Actio...

Page 5: ...e from the actual circumstances Photographic illustrations may also differ and are not binding Other Applicable Documents Product Document P 5xx P 6xx P 7xx piezo positioning systems PZ240 Short Instr...

Page 6: ...cope of delivery of the P 725 The P 725 may only be started operated maintained and cleaned by authorized and qualified staff Safety Precautions CAUTION Dangerous voltage and residual charge in piezo...

Page 7: ...h conductor has to be removed temporarily e g for modification reconnect the P 725 to the protective earth conductor before restarting NOTICE Falling down when removing the thread adapters The P 725 c...

Page 8: ...0 m capacitive sensors LEMO connectors for QuickLock adapters P 725 4CL PIFOC piezo nanofocusing system for long travel ranges 400 m capacitive sensors LEMO connectors for QuickLock adapters P 725 1CA...

Page 9: ...el Product Labeling Labeling Description P 725 2CD Product name example the characters after the period refer to the model 123456789 Serial number example individual for each P 725 Meaning of each pos...

Page 10: ...r Connector P 725 xCD P 725 1CDE2 P 725 xCA P 725 CDD Capacitive sensor D sub E 621 controller module E 625 servo controller benchtop device E 665 servo controller with digital interface benchtop devi...

Page 11: ...pter set M26 1 36 for objective and microscope P 721 08Q QuickLock thread adapter set M19 0 75 for objective and microscope P 721 11Q QuickLock thread adapter set M25 0 75 for objective and microscope...

Page 12: ...pe W0 8 1 36 P 725 13M Thread adapter for microscope SM1 1 035 40 P 725 14M Thread adapter for microscope M34 1 P 725 xxL thread adapter Product no Description Suitability P 725 models P 725 02L Threa...

Page 13: ...y according to the contract and the delivery note 3 If any parts are damaged or missing contact our customer service department immediately p 30 4 Keep all packaging materials in case the product need...

Page 14: ...arth conductor to the protective earth connector of the P 725 as shown in the profile view 3 Tighten the M4 screw with a torque of 1 2 Nm to 1 5 Nm 4 Make sure that the contact resistance at all point...

Page 15: ...t Overview Fig 3 P 721 xxQ thread adapter with PIFOC positioner and microscope objective 1 Thread adapter microscope side 1a Thread for microscope 1b Clamping ring 2 PIFOC positioner 3 Tightening tool...

Page 16: ...f the thread adapter set 1 Narrow side of the key tool 2 Wide side of the key tool Mounting the P 721 xxQ thread adapter set NOTICE Possible damage due to excessive torque Only moderate torque is requ...

Page 17: ...washer on the microscope side is neither too tight nor loose If necessary set the tension of the spring washer by turning the clamping ring Only unscrew the clamping ring as far as necessary 6 Push t...

Page 18: ...e www pi ws P 721 xxA Thread Adapter Set for P 721 xxA and P 725 xCA The P 721 xxA QuickLock thread adapters with large apertures are available with various different inner diameters QuickLock thread...

Page 19: ...et s scope of delivery 3 PIFOC positioner 4 Key tool included in the thread adapter set s scope of delivery 4a Narrow side of the key tool 4b Wide side of the key tool 5 Thread adapter objective side...

Page 20: ...et p 11 Hex key supplied see Fig 6 p 19 Key tool supplied see Fig 6 p 19 Mounting the P 721 xxA thread adapter set 1 Screw the thread adapter for the objective side into the aperture of the PIFOC posi...

Page 21: ...Insert the PIFOC positioner into the thread adapter screwed on the microscope 7 Tighten the thread adapter on the microscope side in the PIFOC positioner using the hex key supplied as shown in the fol...

Page 22: ...Aperture diameter P 725 02M P 725 02L M26 0 75 22 P 725 03M P 725 03L M27 0 75 22 P 725 04M P 725 04L M28 0 75 22 P 725 05M P 725 05L M32 0 75 25 P 725 06M P 725 06L M26 1 36 22 P 725 08M P 725 08L M...

Page 23: ...P 725 xxM thread adapter Mounting the P 725 xxM and P 725 xxL thread adapters 1 Insert the thread adapter for the objective side into the aperture of the PIFOC positioner from below 2 Use the Torx wr...

Page 24: ...INFORMATION Note that only the thread adapter on the objective side is required when mounting the P 725 using the mounting holes in the base body The following figures serve as examples and can diffe...

Page 25: ...8 P 725 xxL thread adapter Clamp the thread adapter in the aperture of the P 725 For details refer to P 725 xxM and P 725 xxL Thread Adapters for P 725 1CDE2 p 22 2 Optional Align the P 725 to the mic...

Page 26: ...on reconnect the P 725 to the protective earth conductor before restarting NOTICE Destruction of the piezo actuator due to electric flashovers The use of the P 725 in environments that increase the el...

Page 27: ...itioner without ID chip The positioner s ID chip does not contain any calibration data Connecting the positioner via LEMO connectors no ID chip lines If the positioning accuracy is reduced after repla...

Page 28: ...e P 725 is accidentally pulled out of the electronics during operation Requirements You have read and understood the safety precautions p 6 Discharging a P 725 Connected to the Electronics In closed l...

Page 29: ...onics before cleaning Prevent cleaning fluid from penetrating the housing of the P 725 NOTICE Damage from ultrasonic cleaning Ultrasonic cleaning can damage the P 725 Do not do any ultrasonic cleaning...

Page 30: ...Screw the objective out before removing Hold the P 725 firmly when you loosen the upper thread adapter To demount the objective and the PIFOC positioner reverse the steps as described in Installation...

Page 31: ...ed 680 470 330 230 as P 725 xCL Hz 20 Resonant frequency under load 150 g 290 185 140 120 as P 725 xCL Hz 20 Push pull force capacity in positioning direction 100 20 100 20 100 20 100 20 as P 725 xCL...

Page 32: ...20 Resonant frequency unloaded 1180 1180 Hz 20 Resonant frequency under load 200 g 450 450 Hz 20 Push pull force capacity in positioning direction 100 20 100 20 N Max Drive properties Ceramic type PI...

Page 33: ...he mechanical resonant frequency 2 The heat that is generated by the piezo actuator during dynamic operation limits the value for maximum power consumption Details can be found at the following websit...

Page 34: ...Co KG Auf der Roemerstrasse 1 76228 Karlsruhe Germany Page 34 40 Phone 49 721 4846 0 Fax 49 721 4846 1019 Email info pi de www pi ws Dimensions Note that the decimal places are separated by a comma i...

Page 35: ...8 27 2020 Physik Instrumente PI GmbH Co KG Auf der Roemerstrasse 1 76228 Karlsruhe Germany Page 35 40 Phone 49 721 4846 0 Fax 49 721 4846 1019 Email info pi de www pi ws Fig 11 P 725 xCD xCL Dimension...

Page 36: ...h 8 27 2020 Physik Instrumente PI GmbH Co KG Auf der Roemerstrasse 1 76228 Karlsruhe Germany Page 36 40 Phone 49 721 4846 0 Fax 49 721 4846 1019 Email info pi de www pi ws Fig 12 P 725 xCA dimensions...

Page 37: ...40 Phone 49 721 4846 0 Fax 49 721 4846 1019 Email info pi de www pi ws Fig 13 P 725 xDD dimensions in mm please order the adapter set separately Torque for Stainless Steel Screws A2 70 Screw size Mini...

Page 38: ...above and P 725 x0L below 1 LEMO coaxial connector for capacitive sensor Probe 2 LEMO coaxial connector for capacitive sensor Target 3 LEMO coaxial connector for piezo voltage LEMO coaxial connectors...

Page 39: ...i ws P 725 SDD Fig 16 LEMO connectors for the P 725 SDD 1 LEMO connector 4 pole for the strain gauge sensor 2 LEMO coaxial connector for piezo voltage Sensor connector Fig 17 LEMO connector 4 pole for...

Page 40: ...voltage 4 Not connected 5 Input Target sensor signal movable part of the capacitive sensor The connector shell is connected to the cable shield Old Equipment Disposal In accordance with EU law electri...

Reviews: