User Manual
P725T0010, applies to P-725
CBo/KSch, 8/27/2020
Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany
Page 20 / 40
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email
Requirements
You have read and understood the safety precautions (p. 6).
The P-725 is
not
connected to the electronics.
Tools and accessories
P-721.xxA QuickLock thread adapter set (p. 11)
Hex key supplied (see Fig. 6, p. 19)
Key tool supplied (see Fig. 6, p. 19)
Mounting the P-721.xxA thread adapter set
1.
Screw the thread adapter for the objective side into the aperture of the PIFOC
positioner from below.
2.
Tighten the thread adapter on the objective side using the wide face of the key tool as
shown in the figure below.
3.
Screw the thread adapter on the microscope side into the microscope.
4.
Tighten the thread adapter on the microscope side with the wide side of the key tool.
5.
Loosen the clamping screw of the aperture using the hex key supplied as shown in the
figure below.