Mounting the P-725.xxM and P-725.xxL thread adapters
Requirements
You have read and understood the safety precautions (p. 6).
The P-725 is
not
connected to the electronics.
Tools and accessories
P-725.xxM thread adapter, microscope side (p. 12)
P-725.xxL thread adapter, objective side (p. 12)
Torx wrench supplied
Key tool (in the scope of delivery for the P-725.xxM thread adapter)
Mounting the P-725.xxM and P-725.xxL thread adapters
1.
Insert the thread adapter for the objective side into the aperture of the PIFOC
positioner from below.
2.
Use the Torx wrench to tighten the screw for the thread adapter on the objective side
(see figure).
3.
Screw the thread adapter on the microscope side into the microscope.
4.
Tighten the thread adapter on the microscope side with the wide side of the key tool.
5.
Insert the PIFOC positioner into the thread adapter on the microscope side.
6.
Use the Torx wrench to tighten the screw for the thread adapter on the microscope
side (see figure).
7.
Screw the objective from below into the thread adapter on the objective side of the
PIFOC positioner. Do not exceed the maximum torque of 0.6 Nm.