3.2.
Image Initialization (Calibration)
In order to maximize the high image quality of the system, it is necessary to correct the
non-uniformity in each pixel that is inherent in the image sensor.
The system is equipped with a function to correct the output value for each pixel using the black level
as a standard. Using a uniform black image as the standard, each pixel's black level is set to zero.
This makes it possible to obtain faithful image output for the input light level.
•
In order to obtain faithful image output, execute calibration when the following settings are changed.
■
When the frame rate is changed
■
When the shutter speed is changed
■
When the resolution is changed
Also, depending on the settings, phenomena such as the following may occur.
■
Fixed noise like vertical bands appears
■
A portion of the screen is clear, but noise appears in the area around it
These phenomena can be resolved by executing calibration again.
3.2.1. Executing Calibration
Calibration is executed from PFV.
•
For the procedure of performing this function, refer to "Remote Keypad User's Manual" or "Photron
FASTCAM Viewer User's Manual".
31
FASTCAM Mini WX50/100 Hardware Manual
Summary of Contents for FASTCAM Mini WX100
Page 1: ...F FA AS ST TC CA AM M M Mi in ni i W WX X5 50 0 1 10 00 0...
Page 12: ......
Page 26: ...Chapter 1 Overview 14...
Page 40: ...Chapter 2 Setup 28...
Page 73: ...61 FASTCAM Mini WX50 100 Hardware Manual...
Page 76: ...Chapter 5 Product Specifications 5 1 Specifications 5 2 Dimensions 64...
Page 80: ...Chapter 5 Product Specifications 68...
Page 94: ...Chapter 5 Product Specifications 5 2 2 AC Power Supply Unit mm 82...
Page 95: ...83...
Page 96: ...Chapter 6 Warranty Chapter 6 Warranty 6 1 About the Warranty 84...
Page 98: ...Chapter 7 Contacting Photron Chapter 7 Contacting Photron 7 1 Contact Information 86...