ACCUCONTROL
Operating notes
59245(e)
39
7.5.3
Charging the ACCU AC 4.5DC directly at the patient lifter
The ACCU AC 4.5DC can only be charged with the CONTROL AC 4.5DC!
No motion functions are possible during the charging!
Figure 21
Charging the ACCU AC 4.5DC at the CONTROL AC 4.5DC
A
PLUG-IN CHARGER PB 4.5
B
Y-cable (for system with 2 motors)
C
ACCU AC 4.5DC
D
CONTROL AC 4.5DC
E
Connecting socket for the adapter cable or
the Y-cable (refer to Figure 5)
F
Patient lifter
G
Adapter cable (for system with 1 motor)
1
Use the adapter cable or the Y-cable to connect the PLUG-IN CHARGER PB 4.5 to the
CONTROL AC 4.5DC (as shown in Figure 21).
2
Plug the PLUG-IN CHARGER PB 4.5 into the outlet. The charge level of the ACCU AC 4.5DC is
indicated by the LED on the PLUG-IN CHARGER PB 4.5 as follows:
- LED is flashing green: the ACCU AC 4.5DC is charging,
- LED is steady green: the ACCU AC 4.5DC is fully charged.
3
When the LED on the PLUG-IN CHARGER PB 4.5 is steady green,
- pull the PLUG-IN CHARGER PB 4.5 out of the outlet and
- unplug the PLUG-IN CHARGER PB 4.5 cable from the ACCU AC 4.5DC.
4
Allow the ACCU AC 4.5DC to rest for about 1 hour after charging it; this will prolong its lifespan.
C
D
E
G
A
B
F