ACCUCONTROL
Technical specifications
59245(e)
17
5. Technical specifications
CONTROL AC 4.5DC
1)
/
CONTROL AC 4.5 BAS
and
PRO
models
Input voltage
24 V DC
Input current for the "BAS" model
Max. 7.0 A DC
Input current for the "PRO" model
Max. 8.5 A DC
Mode of operations
2)
Intermittent duty 2 min./18 min.
Permitted power consumption
Max. 8.5 A DC
Power limit (over-current shutdown)
Adjustable to 8.5 A DC
Audible alarm threshold
21 V DC
Shutdown threshold discharge
17 V DC
ACCU AC 4.5DC
1)
/
ACCU AC 4.5
Rated voltage
24 V DC
Capacity
4.5 Ah
Fuse
7.0 A (PolySwitch)
Battery type
Lead rechargeable battery (Pb)
Self-discharging
After approx. 6 months
Charging time
Approx. 8 hours
ACCUCONTROL 4.5DC
1)
/
ACCUCONTROL 4.5
Protection degree
3)
IP54
1)
DC
= Direct Charging
2)
Mode of operation: intermittent duty 2 min./18 min. This means that after the unit is operated with its
rated load for up to two minutes it must then be paused for 18 minutes. The system can malfunction if
this pause is not observed!
3)
The
degree of protection (IP54) is only met when the "ACCU" battery unit is properly plugged into the
"CONTROL" control unit.