Page 14
5.3.1.5 Correction value
Entry of a correction (offset) value, by which the measured value will be adjusted. The
value will be added to the measured value. Positive as well as negative correction values
are possible. The entry is always carried out to one decimal point.
Menu <–> Select
Set digit,
Select digit
Assuming the measured value is 28.45 and the correction value is –1.5, the display
will indicate 26.95 .
5.3.2 Resistance Measurement
Menu <–> Selection
press key
to accept the selection
Also suitable for non-linear resistances
5.3.2 1 Select Measuring Mode
Menu <–> Selection
2 wire
3 wire
4 wire
4. electrical connections
9
➱
press key
to accept the selection
press key
to accept the selection
Page 15
5.3.2 2 Measurement range
Menu <–> Selection
0 ... 400
Ω
5.3.2.5 Decimal point
16
0 ... 4 k
Ω
5.3.2.5 Decimal point
16
Pt 100 (with stored scaling slope)
5.3.2.3 Display unit
15
Pt 1000 (with stored scaling slope)
5.3.2.3 Display unit
15
➱
➱
➱
➱
5.3.2 3 Display Unit (Pt100, Pt1000)
Menu <–> Selection
Display in °C
Display °F
5.3.2.4 Correction value (Pt100, Pt1000)
Entry of a correction (offset) value, by which the measured value will be adjusted. The
value will be added to the measured value. Positive as well as negative correction values
are possible. The entry is always carried out to one decimal point.
Menu <–> Selection
Set digit,
Select digit
Assuming the measured value is 28.45 and the correction value is –1.5,
the display indicates 26.95.
english
press key
to accept the selection
press key
to accept the selection
press key
to accept the selection