6
Specifications
Warning
: The technical specifications of the mask are provided for the healthcare professional to determine
if it is compatible with the CPAP or bi-level therapy device. If used outside these specifications, or if used
with incompatible devices, the mask may be uncomfortable, the seal of the mask may not be effective,
optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect device function.
Pressure Flow Curve
0
10
20
30
40
50
60
0
5
10
15
20
25
30
Flo
w
R
at
e
(S
LP
M
)
Pressure cm H
2
O
Resistance
Drop in Pressure at
50 SLPM
100 SLPM
All sizes
0.8 cm H
2
O
3.2 cm H
2
O
Deadspace Volume
P
37 mL
S/M
49 mL
L
73 mL
XL (adult only)
85 mL
Sound Levels
A-weighted Sound Power Level 27 dBA
A-weighted Sound Pressure Level at 1 m 19 dBA
Disposal
Dispose of in accordance with local regulations.
Storage Conditions
Temperature: -4º to 140º F (-20º to +60º C)
Relative Humidity: 15% to 95% non-condensing
Summary of Contents for Respironics Wisp Youth