Replacing Batteries
3-8
Maintenance
The display will then show
briefly, and the
Step Off
light will
flash slowly.
Step 7.
Replace the display as follows:
– Place the connection cord through the display cutout.
– Hold the scale display above the top of the column with the
display facing you.
– Plug the cord into the adapter inside the top of the column.
– Center the display over the top of the column so that the plastic
fastener strips on the display and the column match.
– Press down on the display. The back of the display should rest
against the rim on the top of the column.
When the display is properly connected to the Scale:
– the
Step Off
light stops flashing.
– the Scale will say
“zero pounds”,
display
0.0
, and then turn off.
Note
Important:
Do not step on the Scale while doing this step.
If the
Step Off
light does not stop flashing, check the cord connections
by unplugging the cord and replugging it in until it snaps into place.
Also check that the batteries are inserted the right way.
Step 8.
Return the Scale to its original spot.
Caution
Do not lift the Steady Scale.
If the Steady Scale must be moved, ask for assistance or call your
healthcare provider for help.
Step 9.
Perform the
Radio Test
described in the
Troubleshooting
section.
Step 10.
Step on the Scale and verify that it is working properly.
Note
If the Scale does not seem to be operating properly or the
Radio Test
fails, contact your healthcare provider.