Criteria D: Loss of function or degradation of performance which is not recoverable, owning to damage to
hardware or software, or loss of data.
8.4.2
Mains Drop-out Immunity
Standard Reference: IEC 61000-4-11:2004
-
Variation 0% (= 100% dip) at T-event = 50msec. Performance criteria B
-
Variation 40% (= 60% dip) at T-event = 100msec. Performance criteria B
-
Variation 0% (= 100% dip) at T-event = 1 sec. Performance criteria C
Additional for USA:
-
Variation 0% (= 100% dip) at T-event = 100msec in standby mode. Performance criteria B
Performance criteria:
Criteria A: Normal performance;
Criteria B: Temporary loss of function or degradation of performance which ceases after the disturbance
ceases, and from which the equipment under test recovers its normal performance, without operator
intervention;
Criteria C: Temporary loss of function or degradation of performance, the correction of which requires operator
intervention;
Criteria D: Loss of function or degradation of performance which is not recoverable, owning to damage to
hardware or software, or loss of data.
8.4.3
Surge Immunity (Lightning Test)
Standard Reference: IEC61000-4-5:2005
The product shall withstand mains interferences of
For Europe:
Differential Mode: ±2kV / 2ohm, Performance criteria C
Parameters: Bi-wave, Open circuit voltage: 1.2/50µs, Short circuit current: 8/20µs.
Range: ± 0.5kV to ± 2kV, in steps of 0.5kV. 10 shots per combination. One shot per minute.
Serial impedance: 2ohm. Polarity: Positive and Negative. Phase angle: 90
°
and 270
°.
Common Mode: ±6kV / 2ohm, Performance criteria C
Parameters: Bi-wave, Open circuit voltage: 1.2/50µs, Short circuit current: 8/20µs.
Range: ±3kV to ±6kV, in step of 1kV. 10 shots per combination. One shot per minute.
Serial impedance: 2ohm. Polarity: Positive and Negative. Phase angle: 90
°
and 270
°.
For NAFTA:
Differential Mode: ±6kV / 12ohm, Performance criteria C
Parameters: Bi-wave, Open circuit voltage: 1.2/50µs, Short circuit current: 8/20µs.
Range: ± 1kV to ± 6kV in steps of 1kV. 10 shots per combination. One shot per minute.
Serial impedance: 12ohm. Polarity: Positive and Negative. Phase angle: 90
°
and 270
°.
Common Mode: ±6kV / 12ohm, Performance criteria C
Parameters: Ring-wave, 100kHz,
Range: ±3kV to ±6kV, in step of 1.5kV. 10 shots per combination. One shot per minute.
Serial impedance: 12ohm. Polarity: Positive and Negative. Phase angle: 90
°
and 270
°.
For others:
Follow Europe requirements.
With the following requirement,
•
Apparatus should fulfill the leakage current requirements of less than 3mA after the Surge Immunity
test.
•
Defects or permanent deviations are not allowed.
Performance criteria:
Criteria A: Normal performance;
Criteria B: Temporary loss of function or temporary degradation of performance not requiring an operator;
Criteria C: Temporary loss of function or temporary degradation of performance requiring an operator;
Criteria D: Loss of function with permanent damage to equipment (which means failing the test).
2-18