150S4 LCD
37
Go to cover page
Colour adjustment
ZERO CALIBRATION
Zero calibration is performed to determine the output of the measuring probe when no light reaches the sensor and to set this as the zero
point to which all other measurements are referenced. Zero calibration must be performed after the POWER switch has been set ON
brfore taking any measurements.
To perform zero calibration :
* Before performing zero calibration, check that the measuring probe has been connected to probe
connector P1.
1. Check that the POWER switch is set to ON.
2. Set the measuring/viewing selector to the
(viewing) position. (An image can be seen in the
viewfinder, but no light will reach the sensor.)
3. Press 0-CAL.
* If zero calibration is being performed immediately
after the POWER switch has been set to ON, press
0-CAL after "PUSH 0-CAL KEY" appears in the
liquid crystal display.
4. Set the measuring/viewing selector to the
position.
Measurements will be started immediately.
Display after POWER
switch is set to ON
Press 0-CAL,
Performing zero
calibration
Zero calibration
completed
Approximately 5 seconds
DARKEN PROBE
DARKEN PROBE
PUSH 0-CAL KEY
ZERO CALIBRATION
-->
-->
-->
CH00
P1
E1 [
]
Measuring/viewing selector
1. POWER switch
3. 0-CAL
Index
Summary of Contents for 150S4FG/00
Page 15: ...150S4 LCD 15 Go to cover page Wiring Diagram ...
Page 36: ...36 150S4 LCD Go to cover page Colour Adjustment ...
Page 44: ...Block Diagram DSUB DVI 44 150S4 LCD 9 Go to cover page ...
Page 45: ...Block diagram ROM 45 9 Go to cover page 150S4 LCD ...
Page 46: ...Block Diagram Scaler 46 150S4 LCD 9 Go to cover page ...
Page 47: ...Block diagram Power 47 9 Go to cover page 150S4 LCD ...
Page 48: ...Block Diagram Power board 48 150S4 LCD 9 Go to cover page ...
Page 49: ...Block diagram Power board_B 49 9 Go to cover page 150S4 LCD ...
Page 51: ...Power Board C B A 51 9 Go to cover page 150S4 LCD ...
Page 52: ...Block Diagram Scaler C B A 52 150S4 LCD 9 Go to cover page ...
Page 53: ...Solder side C B A 53 9 Go to cover page 150S4 LCD ...