41
4.
Select the type of log data to be exported.
• To export all types of data, press “All Ch”.
• To export only selected data, press the variable to
be exported:
• Actual Temp.
• Actual CO
2
Level
• Door Opening*
Then press “Selected Ch”.
* When both of the Auto-lock function and the User-ID mode are ON (refer to pages 55–57), the User-IDs
entered to unlock the outer door are also exported.
Note:
When the specified operation log data does
not exist, a warning notice is displayed. Press “OK”,
and select a new period, following procedures
3
and
4
.
5.
When export is completed, press “OK” and
remove the USB memory device from the USB
port.
Note:
Even after export is completed, log data
saved in the incubator are not deleted.
Notes:
• A log folder is created in the USB memory device, and the exported file is saved in this folder in CSV
format. The name of the exported file is in the format
date
_
type of data
(eg
20161001-20170101_AllCh.csv or 20170101_Temp.csv).
• On the beginning of the exported file, “MCO-230AIC,” and Unique ID (if registered, refer to page 37) are
written (eg MCO-230AIC, RoomA001).