background image

24

The virtual 3-D model can not be moved.

The endoscopic mode is enabled.

Change to the Classic mode in the view

panel options.

The Patient calibration is not correct.

Recalibrate the Patient in the Setup

menu.

An other additional tracker is in the 

line-of-sight of the camera.

Remove the additional tracker from the 

navigation camera line of sight.

The navigation system is not functioning 

or is not functioning properly.

The camera lens is covered with debris.

Clean the camera lens.

The software application locks up.

The application is experiencing an 

internal malfunction.

Press the CTRL/ALT/Delete buttons simultane-

ously to perform a soft reboot. Press the ON/

OFF button to perform a hard reboot.

PROBLEM

CAUSE

ACTION

An unobstructed line-of-sight does not 

exist between the navigation camera 

lens and the tracker.

Make sure the line-of-sight between the

navigation camera lens and the tracker is

unobstructed.

The pointer, drill, or instrument is out of 

calibration.

Recalibrate the pointer, drill, or instru-

ment. See the To Perform Tool Calibration 

section.

Recalibrate the Patient.

The navigation system is not accurate 

based on visual indications.

The Patient is out of calibration.

The trainer is not detecting critical risk

structures.

The ground cable with magnet is not

installed on the instrument tracker or 

is not making contact with the manual

instrument body.

Make sure the ground cable with magnet

is installed properly.

The ground cable with magnet is not

installed on the skull and positioned into

the allocated cavity on top of the skull.

Make sure the ground cable with magnet

is installed properly.

A tracker is within the navigation camera

line of sight.

Remove the tracker from the navigation

camera line of sight.

The slice selection slider bar does not

work.

The instrument does not appear in the

software.

The pattern on the instrument tracker 

is not in the line-of-sight of the 

camera.

Hold the tracker on the instrument into

the line-of-sight of the camera, while the

pattern on top of the tracker is facing to

the camera.

The camera does not detect the tracker

on the instrument, although the pattern is

facing to the camera.

The detection of the tracker can be

disturbed by very bright light shining 

on the camera or on the pattern of the

tracker.

Make sure the light shining on the

tracker or camera is reduced.

The virtual 3-D model does not appear in

the 3-D-view.

During the endoscopic mode:

An other additional instrument tracker 

is in the line-of-sight of the camera.

Remove the additional instrument tracker

from the line-of-sight of the camera.

TROUBLESHOOTING

Summary of Contents for SINUS TRAINER

Page 1: ...USER MANUAL PHACON SINUS TRAINER...

Page 2: ......

Page 3: ...your trackers your tracking camera your patient your pointer your holder tray with skull your PHACON navigation software DISCOVER THE INTRODUCTION IN OUR PHACON TRAINER...

Page 4: ...simulation 15 Select the axis view 15 Adjust the quality or perspective of the X Y and Z windowpane display 16 Adjust the perspective of the 3 D windowpane display 16 Change the view mode of the 3 D...

Page 5: ...kull 23 Clean the skull holder with tray 23 Clean the trainer accessories 24 TROUBLESHOOTING 24 Troubleshooting 26 Error Messages 27 HOW TO PACK THE TRANSPORT CASE 28 PRODUCT WARRANTIES 29 MANUFACTURE...

Page 6: ...camera F Sinus skull G Holder Tray with adjustable skull fixation H Ground cable 3x I Instrument tracker for ball probe J Pointer K Instrument tracker for suction M 3 x Instrument tracker for differe...

Page 7: ...e image within the navigation camera s line of sight D Simulation Time Used to display the start time of the simulation the current time and the duration of the simulation session E Windowpanes 4 Used...

Page 8: ...al Add a graph in the result screen Select or toggle the language for the software German English Reset the default settings for brightness contrast and slice position in the 4 windowpane in the simul...

Page 9: ...new user click on the CREATE NEW USER button A CREATE NEW USER dialog box will appear Enter the required information into each field and then click OK 3 After a successful LOGIN an existing group can...

Page 10: ...utton to perform instrument and patient calibration for optimal navigation results 6 See To Perform Setup for Navigation section 7 Click on the START button to begin simulation without performing cali...

Page 11: ...rument click on icon Tool calibration RECALIBRATE Fig 5 Fig 5 ALWAYS make sure the round shapes of the pointer drill tracker or instrument tracker are aligned and visible to the navigation camera s li...

Page 12: ...cedure 1 Click on the RECALIBRATE button to begin the calibration process Fig 6 2 Hold the tip of the pointer on the first calibration point of the Patient as indicated in the 3 D windowpane 3 After v...

Page 13: ...egin the calibration process Fig 5 page 11 2 Make sure the desired tool is selected for calibration 3 Hold the tip of the tool in a fixed location on the Patient Make sure the tracking shapes are faci...

Page 14: ...utton to access the VIEW panel options including AXIS VIEW CONTROL VIEW MODE and the OPACITYVALUE of an ANATOMICALSTRUCTURE Fig 8 NOTE The cross hairs visible within each windowpane represent the loca...

Page 15: ...the mouse up and down To adjust the contrast click on the desired windowpane hold down the left mouse button and move the mouse left and right To adjust the zoom click on the desired windowpane hold d...

Page 16: ...t key and left mouse button and move the mouse up and down or left and right The middle mouse button may also be used Fig 10 CHANGE THE VIEW MODE OF THE 3 D WINDOWPANE DISPLAY NOTE The ENDOSCOPIC view...

Page 17: ...erformed to show the tool path during the simulation To virtually extend the tool in the 3 D windowpne click on the EXTEND TOOL button and move the slider to variable extend the virtual tool extension...

Page 18: ...etter visualization within and behind the Patient 1 Click on the drop down list and select the desired anatomical structure Fig 12 2 Move the slider to change the opacity of the selected structure or...

Page 19: ...to see which risk structure was injured and the time the injury occurred Fig 13 The start time current time and du ration of the training session is also displayed END THE SIMULATION OR TRAINING SESSI...

Page 20: ...GRAPH or TABLE button to display the training session data in a graphical or tabular for mat respectively Fig 15 3 Click on the MEASUREMENT drop down list to select DURATION MINUTES or NUMBER OF INJU...

Page 21: ...will allow access to the ANATOMICAL STRUCTURE drop down list Fig 16 4 Click on the ANATOMICAL STRUCTURE drop down list 5 Select the ALL option or a specific risk structure to view the associated numbe...

Page 22: ...the GROUP drop down list and select a group 3 Click on the STATISTICAL PARAMETER drop down list and select a statistical parameter NOTE The STATISTICAL PARAMETER drop down list includes maximum minimu...

Page 23: ...hin the skull using a dry brush and an external vacuum source Make sure the spillways are clear to allow water to low freely during the use of irrigation 5 Wipe all the surfaces of the skull including...

Page 24: ...er is not detecting critical risk structures The ground cable with magnet is not installed on the instrument tracker or is not making contact with the manual instrument body Make sure the ground cable...

Page 25: ...menu with a right click on your desktop and choose NVIDIA Control Panel 2 Select the Manage 3D settings in the left column of the NVIDIA Control Panel 3 In the Global Settings klick on High performan...

Page 26: ...characters The entered group name already exists in the trainer Please enter a unique group name The entered group name is already used Enter a new group name into the trainer maximum of 150 characte...

Page 27: ...and position the two camera arms into the recess next to the Tray the lenses facing down wards Fig 1 2 Cover the camera with the second layer of foam containing a recess for the tracker box Fig 2 3 In...

Page 28: ...s if Products are used with accessories not manufactured by PHACON or if repairs are performed by any party other than authorized PHACON repair personnel Except as set forth above phacon makes no warr...

Page 29: ...29...

Page 30: ...30...

Page 31: ...31...

Page 32: ...PHACON GmbH Karl Heine Str 15 04229 Leipzig Germany Tel 49 0 3 41 47 83 97 32 Fax 49 0 3 41 47 83 91 74 www phacon de...

Reviews:

Related manuals for SINUS TRAINER