2
EN
EN
Range of application
The units TPG 361 respectively TPG 362 must be installed and operated in the following
ambient conditions:
Installation, Setup
The unit can be operated in three different ways:
– mounted in a 19" rack cabinet
– mounted in an operator panel
– as a desktop unit
Rack installation
The unit is designed for installation into a 19" rack chassis adapter according to DIN 41
494. For this purpose, four collar screws and plastic sleeves are supplied with it.
Installation location
weather protected (indoor)
Protection class
IP 20
Installation altitude
max. 2000 m
Ambient temperature
+5°C to +50°C
Relative humidity
5 ... 85 %, non-condensing
Atmospheric pressure
860 hPa - 1060 hPa
DANGER
Commissioning of a defective product
Putting a product which presents a visible damage into operation can be extremely haz-
ardous.
If the product presents a visible damage do not put it into operation and make sure it
is not inadvertently put into operation.
NOTICE
Loss of the protection class of the rack
If the product is installed in a rack, it is likely to lower the protection class of the rack
(protection against foreign bodies and water) e.g. according to the EN 60204-1 regula-
tions for switching cabinets.
Take appropriate measures for the rack to meet the specifications of the protection
class.
NOTICE
Thermal overload!
Set up the unit so that the admissible maximum ambient temperature (e. g. due to
installation or sunlight) is not exceeded.
Ensure adequate air circulation!
Summary of Contents for TPG 36 Series
Page 36: ...36 ZH DIN 41494 19 26 TPG 36x TPG 36x TPG 36x EN 60204 1 1 2 3 EN 60204 1...
Page 37: ...ZH 37 27 28 TPG 36x TPG 36x M3 TPG 36x M3 oder 3 5 122 5 113 70 55 9...
Page 38: ...38 ZH TPG 36x 100 240 V AC 29 S1 0 PE I S1 10 Amax...
Page 42: ...42 KO DIN 41494 19 4 31 TPG 36x TPG 36x TPG 36x EN 60204 1 1 2 3 EN 60204 1...
Page 43: ...KO 43 32 33 TPG 36x TPG 36x 4 M3 TPG 36x 2 M3 oder 3 5 122 5 113 70 55 9...
Page 44: ...44 KO TPG 36x 100 240 V AC 34 S1 0 S1 10 Amax...