EN
7
7-pin C91B appliance con-
nector
Table 1: At the TPG 361 Pin 1 and Pin 6 are not used
Relay output
This connector allows to use the floating switching contacts for an external control sys-
tem.
Pin assignment of the fe-
male 15-pin D-Sub appli-
ance connector:
Table 2: At the TPG 361 Pin 9 and Pin 14 are not used
RS-485
The RS-485 interface allows operation of the TPG 36x via a computer or a terminal. The
use of a Y-connector allows integration into a bus system.
Connect the serial interface to the
RS-485
connector on the rear of the unit using a
shielded cable (electromagnetic compatibility).
5
2
7
3
6
4
1
Pin
Signal
2
Analog output gauge 1 (0 ... 10 V DC)
1
Analog output gauge 2 (0 ... 10 V DC)
5
Screening GND
4
Gauge 1
on: signal
≤
+ 0.8 V DC
off: 2.0 ... 5 V DC or input open
6
Gauge 2
on: signal
≤
+ 0.8 V DC
off: 2.0 ... 5 V DC or input open
3-7
No error
3-7
Error or power supply turned off
NOTICE
Damage to the product
Only connect cables when de-energized.
Never establish a connection using a live cable.
15
9
8
1
Pin
Signal
4
3
2
Switching function 1
Pressure below threshold
Pressure above threshold or power supply turned off
7
6
5
Switching function 2
Pressure below threshold
Pressure above threshold or power supply turned off
11
10
9
Switching function 3
Pressure below threshold
Pressure above threshold or power supply turned off
14
13
12
Switching function 4
Pressure below threshold
Pressure above threshold or power supply turned off
Pin
Supply for relays with higher switching power
15
1
8
+24 V DC, 200 mA
GND
GND
Fuse-protected at 300 mA with PTC element, self-reset-
ting after power off or pulling the relay connector. Meets
the requirements of a grounded protective extra low volt-
age (SELV).
Summary of Contents for TPG 36 Series
Page 36: ...36 ZH DIN 41494 19 26 TPG 36x TPG 36x TPG 36x EN 60204 1 1 2 3 EN 60204 1...
Page 37: ...ZH 37 27 28 TPG 36x TPG 36x M3 TPG 36x M3 oder 3 5 122 5 113 70 55 9...
Page 38: ...38 ZH TPG 36x 100 240 V AC 29 S1 0 PE I S1 10 Amax...
Page 42: ...42 KO DIN 41494 19 4 31 TPG 36x TPG 36x TPG 36x EN 60204 1 1 2 3 EN 60204 1...
Page 43: ...KO 43 32 33 TPG 36x TPG 36x 4 M3 TPG 36x 2 M3 oder 3 5 122 5 113 70 55 9...
Page 44: ...44 KO TPG 36x 100 240 V AC 34 S1 0 S1 10 Amax...