Guide rail
Fig. 1:
Mounting the TPG 36x into the rack frame using a rack chassis adapter
In order to reduce the mechanical strain on the front panel of the TPG 36x, preferably
equip the rack chassis adapter with a guide rail.
Secure the rack adapter in the rack frame.
Slide the TPG 36x into the rack chassis adapter and fasten the adapter panel to the
rack chassis adapter using the screws supplied with the TPG 36x.
Installation in a Con-
trol Panel
1
2
3
NOTICE
Loss of the protection class of the rack
If the product is installed in a rack, it is likely to lower the protection class of the rack
(protection against foreign bodies and water) e.g. according to the EN 60204-1 regula-
tions for switching cabinets.
Take appropriate measures for the rack to meet the specifications of the protection
class.
NOTICE
Thermal overload!
Set up the unit so that the admissible maximum ambient temperature (e. g. due to
installation or sunlight) is not exceeded.
Ensure adequate air circulation!