Installation
Fig. 5:
Assemble connection cable
Pin assignment
Solder in cable according pin assignment.
Assemble cable socket.
Connect the other end of the transmitter cable according to the pin assignment of the
appropriate control device.
Connect the transmitter to the control unit using the connection cable.
Switch on the control unit.
NOTICE
Damaging the transmitter
Incorrect connection, incorrect polarity or inadmissible supply voltages can damage the
transmitter.
Ground (pin 5) and shielding (pin 6) must always be grounded at the controller.
2
5
3
1
6
4
Pin
Pin assignment
1
Identification (3.0 k
Ω
to GND)
2
Analog output 0 ... 10V, (measuring range 1.5 ... 8.5V)
3
Analog ground
4
Supply (V DC), 15 ... 30V
5
Ground (GND)
6
Shielding