16
Technical data and dimensions
11.3 Dimensions
11.4 Conversion
Conversion formula
p=10
(U-c)
<=> U=c+log
10
p
●
p: Pressure
●
U: Measuring signal
●
c: Constant
valid in the range: 1 • 10
-4
hPa < p < 1000 hPa
Conversion curves
ø 32
62
20
20,5
DN 16 ISO-KF
ø 30
VACUUM
DN 16 CF-F
ø 32
74
20
32,3
ø 30
VACUUM
U
p
c
[V]
[mbar, hPa]
5.5
[V]
[Torr]
5.625
[V]
[mTorr]
2.625
[V]
[micron]
2.625
[V]
[Pa]
3.5
[V]
[kPa]
6.5
1E
+05
1E
+04
1E
+03
1E
+02
1E
+01
1E
+00
1E
–01
1E
–02
1E
–03
1E
–04
1E
–05
ov
er
ra
n
g
e
Pa
hPa
Torr
0.0 0.5 1.0 1.5
2.5
3.5
4.5
5.5
6.5
7.5
8.5
9.5
10.3
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
s
enso
r err
or
un
d
e
rr
an
ge
p
U [V]