Service
18
9.2 Re-starting
Check pumping station for contamination and moisture.
If necessary, have Pfeiffer Vacuum Service clean the pumping station completely.
Installation and commissioning in accordance with the operating instructions.
9.3 Disposal
Products or parts thereof (mechanical and electrical components, operating fluids, etc.)
may cause environmental burden.
Safely dispose of the materials according to the locally applicable regulations.
10 Service
Pfeiffer Vacuum offers first-class service!
• Maintenance/repairs on site by Pfeiffer Vacuum field service
• Maintenance/repairs in a nearby service center or service point
• Fast replacement with exchange products in mint condition
• Advice on the most cost-efficient and quickest solution
Detailed information and addresses at:
www.pfeiffer-vacuum.de (Service
).
Maintenance and repairs in Pfeiffer Vacuum ServiceCenter
The following steps are necessary to ensure a fast, smooth servicing process:
Download the forms "Service Request" and "Declaration on Contamination".
1)
Fill out the "Service Request" form and send it by fax or e-mail to your Pfeiffer Vacuum
service address.
Include the confirmation on the service request from Pfeiffer Vacuum with your ship-
ment.
Fill in the contamination declaration and enclose it in the shipment (required!).
Dismantle all accessories.
Drain operating fluid/lubricant.
Drain cooling medium, if used.
Send the pump or unit in its original packaging if possible.
Sending of contaminated pumps or devices
No units will be accepted if they are contaminated with micro-biological, explosive or ra-
dioactive substances. “Hazardous substances” are substances and compounds in ac-
cordance with the hazardous goods directive (current version). If pumps are contaminat-
ed or the declaration on contamination is missing, Pfeiffer Vacuum performs
decontamination at the shipper's expense.
Neutralise the pump by flushing it with nitrogen or dry air.
Close all openings airtight.
Seal the pump or unit in suitable protective film.
NOTE
Note the risk of corrosion for storage
Storage of vacuum pumps can lead to corrosion and aging of lubricants/operating flu-
ids and elastomer seals.
Observe the notes in the operating manuals for the individual components.
Change lubricants and operating fluids before restarting operation.
1)
Forms under www.pfeiffer-vacuum.de