background image

6

3. Installation 

3.1. Setting Up The Pump And Location

Obey national safety regulations and safety requirements
concerning the use of vacuum and electrical equipment.

Remove all packing material, remove the product from its
packing-box, remove the protective covers from the inlet
and outlet ports and keep, inspect the equipment.

If the equipment is not used immediately, replace the
protective covers. Store the equipment in suitable conditi-
ons.

Read and obey this manual before installing or operating
the equipment.

If the equipment is brought from cold environment into a
room for operation, allow the equipment to warm up (pay
attention to water condensation on cold surfaces).

Make sure ventilation is adequate if pump is installed in a
housing or if ambient temperature is elevated.

Provide a firm level platform for the equipment and check
that the system to be evacuated is mechanically stable
and that all fittings are secure.
Attention: Flexible elements tend to shrink when
evacuated.

Adopt suitable measures in case of differences, e. g. using
the equipment outdoors, installation in altitudes of more
than 2000 m above mean sea level, conductive pollution or
dewiness.

Use fixing holes to screw-on the pump to a base plate.

Ambient temperature range + 5 ... +40 °C.

Where rack installation is involved, ensure adequate venti-
lation.

3.2. Connecting The Vacuum Side

Remove locking cap on intake connection.

Make connection between the vacuum system and pump
as short as possible.
– Ensure no mechanical forces act on the pump by a rigid

piping system; use elastic connections or bellows.

Connect pump with intake connection to the apparatus.

Obey maximum permitted pressures and pressure differen-
ces, see section ”Technical data”. Do not operate the
pump with overpressure at the inlet.

If liquid - which would generate vapours - is present in the
system to be evacuated, a condensate trap must be fitted
upstream of the pump.

3.3. Connecting The Pressure Side

Pressure can rise to dangerous levels in exhaust

lines. Therefore, lay exhaust side lines without
shut-off units. Do not connect the exhaust side
with a closed system on account of the danger
of bursting.

In certain applications, exhaust gases and vapours can be
very hot and represent a health and/or environment hazard.

Lay lines from the pump sloping downwards so that
condensate cannot run back into the pump, otherwise fit a
separator.

Due to the high compression ratio of the pumps, pressure
at the outlet port might be generated being higher than the
max. permitted pressure compatible with the mechanical
stability of the system.

Do not permit any uncontrolled pressurizing (e. g. make
sure that the exhaust pipeline cannot become blocked). If
you have an exhaust-isolation valve, make sure that you
cannot operate the equipment with the valve closed. Risk
of bursting!

Ensure that the system design does not allow the exhaust
pipeline to become blocked.

The diameter of the inlet and outlet pipeline should be at
the least as large as the diameter of the pump connection
pipelines.

CAUTION

Summary of Contents for MVP 006-4

Page 1: ...Diaphragm Pump PU 0034 BE B 0606 MVP 006 4 Betriebsanleitung Operating Instructions...

Page 2: ...The Pressure Side 6 3 4 Connecting To Power Supply 7 4 Operations 8 4 1 Important Information 8 4 2 Switching On The Pump 8 4 3 Select rotation speed control 8 4 4 Shutting Down The Pump 8 5 What To...

Page 3: ...instructions in the Service Section Due to the residual leak rate of the equipment there may be an exchange of gas albeit extremely slight between the environment and the vacuum system Adopt suitable...

Page 4: ...Pressure side Power supply connection Position numbers Identical components and accessories parts have the same position numbers in all illustrations Danger of damage to the pump or system CAUTION Da...

Page 5: ...d The pumping of gases which are contaminated with particles dust and condensate The Pump may not be used for the purpose of generating pressure The pumping of liquids Connection to pumps and units wh...

Page 6: ...between the vacuum system and pump as short as possible Ensure no mechanical forces act on the pump by a rigid piping system use elastic connections or bellows Connect pump with intake connection to...

Page 7: ...th pump base to frame potential as defined Stand alone operation with TPS 100 Connect power supply via connection cable PM 051 421 T with TPS 100 The pump starts immediately when the power unit is swi...

Page 8: ...alone operation In the the standalone operation mode the pump can be opera ted in two different modes High speed mode and Low speed mode To select the operating mode the DIP switch 2 on the control bo...

Page 9: ...at atmospheric pressure for condensate a few minutes Deposits have been formed Clean and inspect pump heads inside the pump Hose connectors between pump Check hose connectors between pump are leaky h...

Page 10: ...e dismantling Detach intake and exhaust line from the pump As far as possible place pump on its side so that the head to be dismantled points up Use open ended wrench SW 12 to remove fitting 10 1 at t...

Page 11: ...4 3 3 3 2 2 2 view B Ansicht A Ansicht B B 1 Dismantling the membrane heads 1 Membrane head 1 2 Membrane head 2 3 Membrane head 3 4 Membrane head 4 10 Interhead connection 10 1 Hose fitting 11 Crossh...

Page 12: ...nits which are contaminated microbiologically explo sively or radioactively cannot be accepted as a matter of principle Fill Out The Contamination Declaration In every case the Contamination Declarati...

Page 13: ...ssure mbar 1050 Permissible exhaust pressure mbar 1300 Leak rate mbar l s 6 10 2 Max operating altitude a s l m approx 1000 Ambient temperature range C 5 40 Emission sound pressure level at 1000 min 1...

Page 14: ...ch Come Into Contact With The Medium 1 TPS 100 Power supply unit for MVP 006 4 2 TPS 150 Power supply unit for MVP 006 4 TMH 071 TPD 011 Pos Description Pieces Size Number Comments Order Quantity Oper...

Page 15: ...he repair and or service of vacuum components will only be carried out if a correctly completed declaration has been submitted Non completion will result in delay The manufacturer could refuse to acce...

Page 16: ...WG entspricht We hereby certify that the product specified below is in accordance with the provision of EU Machinery Directive 98 37 EEC EU Electromagentic Compatibility Directive 89 336 EEC and EU Lo...

Page 17: ...17...

Page 18: ...9 0 64 41 8 02 2 02 info pfeiffer vacuum de www pfeiffer vacuum net Vacuum is nothing but everything to us Turbopumps Rotary vane pumps Roots pumps Dry compressing pumps Leak detectors Valves Componen...

Reviews: