1. Safety Precautions 

Obey all relevant safety requirements (regulations and guide-
lines) and adopt suitable safety measures.

Read and follow all the instructions in this manual.

Inform yourself regarding:
- Hazards which can be caused by the pump;
- Hazards which can arise in your system;
- Hazards which can be caused by the medium being

pumped.

Use the equipment for the intended use only (for generation
of vacuum).

Avoid pumping of suddenly decomposing gases.

Adopt suitable measures to prevent any danger arising
from the formation of hot surfaces or electric sparks.

Do not expose any body parts to the vacuum.

Obey notes on correct vacuum and electrical connections,
see section ”Use and operation”.

Make sure that the individual components are only
connected, combined and operated according to their
design and as indicated in the instructions for use.

Observe the safety and accident prevention regulations.

Check regularly that all safety precautions are being
complied with.

Do not carry out any unauthorised conversions or
modifications on the pump.

Check that mains voltage and current conform with the
equipment (see rating plate). 

When returning the pump to us please note the shipping
instructions in the Service Section.

Due to the residual leak rate of the equipment, there may be
an exchange of gas, albeit extremely slight, between the
environment and the vacuum system.

Adopt suitable measures to prevent contamination of the
pumped substances or the environment.

Electronic equipment is never 100% fail-safe. This may lead to
an indefinite status of the equipment. Provide protective
measures against misfunction and failure.

Stand still of the pump or operating a valve must not lead
to a critical dangerous situation under any circumstances.

Ensure that in case of failure the pump and the vacuum
system always will turn into a safe status.

In case of diaphragm cracks or leaks in the manifold
pumped substances mighty be re-leased into the environ-
ment or into the pump housing.

Obey especially notes on operation and use and
maintenance.

Use only genuine spare parts and accessories.

Otherwise safety and performance of the equipment as
well as the electromagnetic compatibility of the equipment
might be reduced.

Ensure that maintenance is done only by suitable trained
and supervised technicians. Obey local and national safety
regulations. If necessary the pump should be
decontaminated.

Before starting maintenance work vent the system, isolate
the pump and other components from the vacuum system
and the electrical supply and allow sufficient cooling of the
pump.

Summary of Contents for MVP 006-4

Page 1: ...Diaphragm Pump PU 0034 BE B 0606 MVP 006 4 Betriebsanleitung Operating Instructions...

Page 2: ...The Pressure Side 6 3 4 Connecting To Power Supply 7 4 Operations 8 4 1 Important Information 8 4 2 Switching On The Pump 8 4 3 Select rotation speed control 8 4 4 Shutting Down The Pump 8 5 What To...

Page 3: ...instructions in the Service Section Due to the residual leak rate of the equipment there may be an exchange of gas albeit extremely slight between the environment and the vacuum system Adopt suitable...

Page 4: ...Pressure side Power supply connection Position numbers Identical components and accessories parts have the same position numbers in all illustrations Danger of damage to the pump or system CAUTION Da...

Page 5: ...d The pumping of gases which are contaminated with particles dust and condensate The Pump may not be used for the purpose of generating pressure The pumping of liquids Connection to pumps and units wh...

Page 6: ...between the vacuum system and pump as short as possible Ensure no mechanical forces act on the pump by a rigid piping system use elastic connections or bellows Connect pump with intake connection to...

Page 7: ...th pump base to frame potential as defined Stand alone operation with TPS 100 Connect power supply via connection cable PM 051 421 T with TPS 100 The pump starts immediately when the power unit is swi...

Page 8: ...alone operation In the the standalone operation mode the pump can be opera ted in two different modes High speed mode and Low speed mode To select the operating mode the DIP switch 2 on the control bo...

Page 9: ...at atmospheric pressure for condensate a few minutes Deposits have been formed Clean and inspect pump heads inside the pump Hose connectors between pump Check hose connectors between pump are leaky h...

Page 10: ...e dismantling Detach intake and exhaust line from the pump As far as possible place pump on its side so that the head to be dismantled points up Use open ended wrench SW 12 to remove fitting 10 1 at t...

Page 11: ...4 3 3 3 2 2 2 view B Ansicht A Ansicht B B 1 Dismantling the membrane heads 1 Membrane head 1 2 Membrane head 2 3 Membrane head 3 4 Membrane head 4 10 Interhead connection 10 1 Hose fitting 11 Crossh...

Page 12: ...nits which are contaminated microbiologically explo sively or radioactively cannot be accepted as a matter of principle Fill Out The Contamination Declaration In every case the Contamination Declarati...

Page 13: ...ssure mbar 1050 Permissible exhaust pressure mbar 1300 Leak rate mbar l s 6 10 2 Max operating altitude a s l m approx 1000 Ambient temperature range C 5 40 Emission sound pressure level at 1000 min 1...

Page 14: ...ch Come Into Contact With The Medium 1 TPS 100 Power supply unit for MVP 006 4 2 TPS 150 Power supply unit for MVP 006 4 TMH 071 TPD 011 Pos Description Pieces Size Number Comments Order Quantity Oper...

Page 15: ...he repair and or service of vacuum components will only be carried out if a correctly completed declaration has been submitted Non completion will result in delay The manufacturer could refuse to acce...

Page 16: ...WG entspricht We hereby certify that the product specified below is in accordance with the provision of EU Machinery Directive 98 37 EEC EU Electromagentic Compatibility Directive 89 336 EEC and EU Lo...

Page 17: ...17...

Page 18: ...9 0 64 41 8 02 2 02 info pfeiffer vacuum de www pfeiffer vacuum net Vacuum is nothing but everything to us Turbopumps Rotary vane pumps Roots pumps Dry compressing pumps Leak detectors Valves Componen...

Reviews: