10 Malfunctions
WARNING
Danger to life from electric shock in the event of a fault
In the event of a fault, devices connected to the mains may be live. There is a danger to life from
electric shock when making contact with live components.
► Always keep the mains connection freely accessible so you can disconnect it at any time.
WARNING
Danger to life from poisoning where toxic process media leak from damaged connections
Sudden twisting of the turbopump in the event of a fault causes fittings to accelerate. There is the risk
of damaging on-site connections (e.g., fore-vacuum line) and resulting leaks. This results in leakage
of process media. In processes involving toxic media, there is a risk of injury and danger to life due to
poisoning.
► Keep masses connected to the turbopump as low as possible.
► Use flexible lines to connect to the turbopump where necessary.
WARNING
Danger to life from the turbopump breaking away in the event of a fault
Sudden jamming of the rotor generates high destructive torques in accordance with ISO 27892. If the
turbopump is
not
properly secured, it can shear off. The energy that this would release could throw
the entire turbopump or shattered pieces from its interior through the surrounding space. Potentially
dangerous gases can escape. There is a risk of very serious injuries, including death, and extensive
property damage.
► Follow the installation instructions for this turbopump.
► Observe the requirements regarding stability and design of the counter flange.
► Use only original accessories or fixing material approved by Pfeiffer Vacuum for the installation.
WARNING
Risk of injury caused by the turbopump breaking away with the vibration compensator in the
event of a malfunction
Sudden jamming of the rotor generates high destructive torques in accordance with ISO 27892.
When using a vibration compensator, this will probably lead to the turbopump being sheared off in
use. The energy that this would release could throw the entire turbopump or shattered pieces from its
interior through the surrounding space. Potentially dangerous gases can escape. There is a risk of
very serious injuries, including death, and extensive property damage.
► Take suitable safety precautions on-site for the compensation of the occurring torques.
► Before installing a vibration compensator, you must first of all contact Pfeiffer Vacuum.
Should malfunctions occur, you can find information about potential causes and how to fix them here.
The operating instructions of the associated electronic drive unit contains more detailed error descrip-
tions.
Problem
Possible causes
Remedy
Turbopump will not start up;
none of the built-in LEDs on the
electronic drive unit light up
● Current supply interrupt-
ed
● Check the plug contacts on the power supply pack.
● Check the current supply lines.
● Check the output voltage at the power supply pack
“DC out” connection.
─ Depending on the version of the power supply
pack, 24 V DC or 48 V DC are present.
● Incorrect operating volt-
age
● Observe the electronic drive unit rating plate.
● Supply the correct operating voltage.
● No operating voltage
present
● Supply the correct operating voltage.
● Switch on the power supply pack.
● Electronic drive unit de-
fective
● Replace the electronic drive unit.
● Contact Pfeiffer Vacuum Service.
Malfunctions
49/62