Problem
Possible causes
Remedy
Pumping speed of pump
too low
● The intake line is not suita-
bly dimensioned
● Make sure that connections are
short and cross sections adequate-
ly dimensioned.
● Exhaust pressure too high
● Check exhaust line outlet opening
and exhaust side accessories.
Loss of operating fluid
● Casing seal is leaking
● Check and replace the seal if nec-
essary.
● Radial shaft seal ring
(RSSR) leaking – not appli-
cable to pumps with mag-
netic coupling
1. Check and replace RSSR if neces-
sary.
2. Check and also replace the associ-
ated bushing if necessary.
● Operational operating fluid
loss – no oil mist filter
(ONF/OME)
● Install ONF/OME.
● Operational operating fluid
loss – with oil mist filter
(ONF/OME)
● Clean ONF/OME and oil return
(ODK); replace filter if necessary.
Unusual noises during
operation
● Silencer is contaminated
● Clean the silencer or replace it.
● Pumping system is conta-
minated or damaged
● Clean and service pump; contact
Pfeiffer Vacuum if necessary.
● Motor bearing is faulty
● Replace motor; contact Pfeiffer
Vacuum if necessary.
Tbl. 11:
Troubleshooting for rotary vane pumps
Malfunctions